DOI 10.1111/j.1540-8159.2007.00715.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2007

Dislodgement of atrial lead of the CRT-D system and its migration into the pocket presenting as its absence on chest x-ray.

Bhargava Kartikeya, Kler Tarlochan Singh

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Lead dislodgement and migration is a known complication after implantation of cardiac rhythm devices. We present a case of dislodgement of atrial lead of the CRT-D system and its migration into the device pocket. The device itself had rotated by 90 degrees but the right and left ventricular leads were not at all affected. The case is unique because the lead had coiled up and was lying behind the device in the pocket and hence was not even seen on a chest x-ray.

Для цитирования: Bhargava Kartikeya, Kler Tarlochan Singh. Dislodgement of atrial lead of the CRT-D system and its migration into the pocket presenting as its absence on chest x-ray.. Pacing and clinical electrophysiology : PACE. 2007. DOI: 10.1111/j.1540-8159.2007.00715.x.