DOI 10.1111/j.1540-8167.2007.00770.x
A prospective comparison of AV delay programming methods for hemodynamic optimization during cardiac resynchronization therapy.
Аннотация
INTRODUCTION
There are several methods for programming the optimal AV delay (AVD) during cardiac resynchronization therapy (CRT). These include Doppler echocardiographic measurements of mitral inflow or aortic outflow velocities, an arbitrarily fixed AVD, and calculations based on intracardiac electrogram (EGM) intervals. The present study was designed to compare the acute effects of AVD programming methods during CRT.
METHODS AND
RESULTS
We studied 28 patients at CRT implant with invasive measurements of LV dP/dt to determine the effect of AVD during atrial sensed (AS) and atrial paced (AP) modes. The optimal AVD, defined as that resulting in the maximal LV dP/dt, was then compared with that predicted by several noninvasive methods. CRT increased LV dP/dt 11% +/- 11% during AS (heart rate: 73 +/- 14 bpm) and 17% +/- 12% during AP (heart rate: 86 +/- 12 bpm) (P < 0.001 vs AS). There was an excellent correlation between the EGM method and the maximum achievable LV dP/dt (AS: R2 = 0.99, P < 0.0001, AP: R2 = 0.96, P < 0.0001) and this method performed better than other techniques.
CONCLUSIONS
An electrogram-based optimization method accurately predicts the optimal AVD among patients over a wide range of QRS intervals during CRT in both AS and AP modes. This simple technique may obviate the need for echocardiography for AVD programming.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.