Pacing and clinical electrophysiology : PACE · 2007
Long-term follow-up of biventricular pacing using a totally endocardial approach in patients with end-stage cardiac failure.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pasquié J L, Massin F, Macia J C, Gervasoni R, Bortone A, Cayla G, Grolleau R, Leclercq F
Аннотация
BACKGROUND: Besides standard left ventricular (LV) stimulation via the coronary sinus, a transseptal approach allows left ventricular endocardial stimulation. We report our long-term observations with biventricular stimulation, using a strictly endocardial system for patients presenting with severe congestive heart failure.
METHODS: Six patients with nonischemic cardiomyopathy (mean age = 60 +/- 9.6 years, women) in New York Heart Association (NYHA) functional class III (n = 5) or IV, despite optimal drug therapy, and a mean LV ejection fraction of 24 +/- 3%, underwent implantation of biventricular stimulation systems between April 1998 and March 1999. All presented with left bundle branch block and an increased LV end-diastolic diameter (mean = 66 +/- 5 mm). In all patients, a bipolar pacing lead was implanted in the lateral LV wall using a direct transseptal approach. After implantation, all patients received oral anticoagulation.
RESULTS: QRS duration decreased from 184 +/- 22 ms to 108 +/- 11 ms. NYHA functional class decreased to II in all patients within 1 month. Over a 85 +/- 5 month follow-up, two patients underwent cardiac transplantation, 2 and 4 years after device implantation, respectively; two patients died of end-stage heart failure 4 years after system implantation; and two patients were alive in functional class II. One patient, who experienced syncope due to fast ventricular, underwent implantation of an ICD. One transient ischemic attack occurred in a patient whose anticoagulation was temporarily interrupted.
CONCLUSIONS: Long-term endocardial biventricular stimulation via a transseptal approach was safe and effective in this small population. This approach needs to be further compared with conventional epicardial pacing via the coronary sinus.
METHODS: Six patients with nonischemic cardiomyopathy (mean age = 60 +/- 9.6 years, women) in New York Heart Association (NYHA) functional class III (n = 5) or IV, despite optimal drug therapy, and a mean LV ejection fraction of 24 +/- 3%, underwent implantation of biventricular stimulation systems between April 1998 and March 1999. All presented with left bundle branch block and an increased LV end-diastolic diameter (mean = 66 +/- 5 mm). In all patients, a bipolar pacing lead was implanted in the lateral LV wall using a direct transseptal approach. After implantation, all patients received oral anticoagulation.
RESULTS: QRS duration decreased from 184 +/- 22 ms to 108 +/- 11 ms. NYHA functional class decreased to II in all patients within 1 month. Over a 85 +/- 5 month follow-up, two patients underwent cardiac transplantation, 2 and 4 years after device implantation, respectively; two patients died of end-stage heart failure 4 years after system implantation; and two patients were alive in functional class II. One patient, who experienced syncope due to fast ventricular, underwent implantation of an ICD. One transient ischemic attack occurred in a patient whose anticoagulation was temporarily interrupted.
CONCLUSIONS: Long-term endocardial biventricular stimulation via a transseptal approach was safe and effective in this small population. This approach needs to be further compared with conventional epicardial pacing via the coronary sinus.