Clinical evaluation of the rapid nucleic acid amplification point-of-care test (Smart Gene SARS-CoV-2) in the analysis of nasopharyngeal and anterior nasal samples.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kiyasu Yoshihiko, Owaku Masato, Akashi Yusaku, Takeuchi Yuto, Narahara Kenji, Mori Sunao, Nagano Takashi, Notake Shigeyuki, Ueda Atsuo, Nakamura Koji, Ishikawa Hiroichi, Suzuki Hiromichi

Аннотация

INTRODUCTION: Smart Gene is a point-of-care (POC)-type automated molecular testing platform that can be performed with 1 min of hands-on-time. Smart Gene SARS-CoV-2 is a newly developed Smart Gene molecular assay for the detection of SARS-CoV-2. The analytical and clinical performance of Smart Gene SARS-CoV-2 has not been evaluated.

METHODS: Nasopharyngeal and anterior nasal samples were prospectively collected from subjects referred to the local PCR center from March 25 to July 5, 2021. Two swabs were simultaneously obtained for the Smart Gene SARS-CoV-2 assay and the reference real-time RT-PCR assay, and the results of Smart Gene SARS-CoV-2 were compared to the reference real-time RT-PCR assay.

RESULTS: Among a total of 1150 samples, 68 of 791 nasopharyngeal samples and 51 of 359 anterior nasal samples were positive for SARS-CoV-2 in the reference real-time RT-PCR assay. In the testing of nasopharyngeal samples, Smart Gene SARS-CoV-2 showed the total, positive and negative concordance of 99.2% (95% confidence interval [CI]: 98.4-99.7%), 94.1% (95% CI: 85.6-98.4%) and 99.7% (95% CI: 99.0-100%), respectively. For anterior nasal samples, Smart Gene SARS-CoV-2 showed the total, positive and negative concordance of 98.9% (95% CI: 97.2-99.7%), 98.0% (95% CI: 89.6-100%) and 99.0% (95% CI: 97.2-99.8%), respectively. In total, 5 samples were positive in the reference real-time RT-PCR assay and negative in the Smart Gene SARS-CoV-2 assay, whereas 5 samples were negative in the reference real-time RT-PCR assay and positive in the Smart Gene SARS-CoV-2 assay.

CONCLUSION: Smart Gene SARS-CoV-2 showed sufficient analytical performance for the detection of SARS-CoV-2 in nasopharyngeal and anterior nasal samples.