International journal of radiation oncology, biology, physics · 2010
RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemoradiotherapy.
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Doornaert Patricia, Verbakel Wilko F A R, Bieker Michael, Slotman Ben J, Senan Suresh
Аннотация
PURPOSE: Volumetric modulated arc therapy (RapidArc, Varian Medical Systems) permits the delivery of highly conformal dose distributions. We studied planning and delivery in patients who underwent RapidArc for locally advanced head-and-neck cancer (HNC).
METHODS AND MATERIALS: A total of 35 consecutive patients who completed RapidArc with concurrent chemotherapy for Stages III-IV tumors of the oro- and hypopharynx/larynx in our center were identified. All underwent bilateral neck irradiation and 21 patients had at least N2 disease. A simultaneous integrated boost (SIB) delivered 70 Gy (in 2 Gy/fraction) to the planning target volume (PTV)(boost) and elective nodal regions (PTV(elect)) received 57.75 Gy. A standard planning constraint set was used and constraints for parotid glands were individually adapted. Treatments were delivered using two arcs after all plans were verified in a solid water phantom using GafChromic External Beam Therapy films.
RESULTS: RapidArc planning generally took 1.5-2 h, which was faster than with our previous seven-field intensity-modulated radiotherapy sliding window technique. Film dosimetry revealed that 0.6% of films exceeded a combination of dose differences ≥3% or distance to agreement ≥2 mm. More than 99% of both PTVs received ≥95% of the prescription dose. Average plan conformity index was 1.13 and mean dose to ipsilateral and contralateral parotid glands were 31.4 Gy and 26.1 Gy, respectively. The mean beam-on time was <3 min and mean number of monitor units was 426.
CONCLUSIONS: RapidArc achieved excellent target coverage and normal tissue sparing, with delivery completed in less than 3 min. RA is currently our standard intensity-modulated radiotherapy approach for advanced HNC.
METHODS AND MATERIALS: A total of 35 consecutive patients who completed RapidArc with concurrent chemotherapy for Stages III-IV tumors of the oro- and hypopharynx/larynx in our center were identified. All underwent bilateral neck irradiation and 21 patients had at least N2 disease. A simultaneous integrated boost (SIB) delivered 70 Gy (in 2 Gy/fraction) to the planning target volume (PTV)(boost) and elective nodal regions (PTV(elect)) received 57.75 Gy. A standard planning constraint set was used and constraints for parotid glands were individually adapted. Treatments were delivered using two arcs after all plans were verified in a solid water phantom using GafChromic External Beam Therapy films.
RESULTS: RapidArc planning generally took 1.5-2 h, which was faster than with our previous seven-field intensity-modulated radiotherapy sliding window technique. Film dosimetry revealed that 0.6% of films exceeded a combination of dose differences ≥3% or distance to agreement ≥2 mm. More than 99% of both PTVs received ≥95% of the prescription dose. Average plan conformity index was 1.13 and mean dose to ipsilateral and contralateral parotid glands were 31.4 Gy and 26.1 Gy, respectively. The mean beam-on time was <3 min and mean number of monitor units was 426.
CONCLUSIONS: RapidArc achieved excellent target coverage and normal tissue sparing, with delivery completed in less than 3 min. RA is currently our standard intensity-modulated radiotherapy approach for advanced HNC.