DOI 10.1177/000348940911800302
Normal values for pharyngeal pH monitoring.
Аннотация
OBJECTIVES
We performed a prospective study of asymptomatic adult volunteers to establish normative values of pharyngeal pH using a novel pH probe.
METHODS
The Dx-pH probe is a novel pH device capable of measuring liquid and aerosolized acid levels. Twenty asymptomatic patients (Reflux Symptom Index less than 10 and Reflux Finding Score less than 6) underwent simultaneous investigation with this probe placed in the oropharynx and a dual antimony probe placed in the hypopharynx and esophagus. The reflux parameters measured from the oropharyngeal probe included the percentage of time and the number of events in which the pH was less than 5.5,5.0,4.5, and 4.0.
RESULTS
The upper limits of normal (95th percentile) for the number of events below pH of 5.5, 5.0, 4.5, and 4.0 per 24-hour period were 16.6, 10.7, 7.4, and 0.2, respectively. The upper limits of normal (95th percentile) for an acid exposure time below pH of 5.5, 5.0, 4.5, and 4.0 per 24-hour period were 820 seconds, 385 seconds, 75 seconds, and 3 seconds, respectively.
CONCLUSIONS
Normative pharyngeal pH values are presented. Further studies are required to determine clinical relevance.