MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcin.2016.03.011

Медицинская библиотека · JACC. Cardiovascular interventions · 2016

Transcatheter Procedure for Residual Mitral Regurgitation After MitraClip Implantation Using Amplatzer Duct Occluder II.

Kubo Shunsuke, Cox Justin M, Mizutani Yukiko, Uberoi Abhimanyu, Chakravarty Tarun, Nakajima Yoshifumi, Hussaini Asma, Tat Emily, Makar Moody, Kar Saibal

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study reports a novel transcatheter procedure for residual mitral regurgitation (MR) after MitraClip implantation using the Amplatzer Duct Occluder II (ADO II).

BACKGROUND

Although the MitraClip procedure is a transcatheter treatment option for patients at high surgical risk with severe MR, management of significant residual MR after MitraClip implantation is still challenging.

METHODS

We describe a case series of 9 consecutive patients who underwent transcatheter deployment of the ADO II plug for significant residual MR after MitraClip implantation from April to October 2015.

RESULTS

The mean age was 79.3 ± 11.4 years. The deployment of the ADO II plug was performed at the initial MitraClip procedure in 7 patients and at the second procedure for recurrent symptoms in 2 patients. There were 2 types of residual MR seen after MitraClip implantation: residual commissural MR (n = 3) and residual intraclip MR (n = 6). The ADO II was successfully deployed with significant reduction of MR flow and left atrial pressure in all patients. The ADO II plug was retrieved in 1 patient because of device embolization to the ostial right coronary artery. However, all patients were discharged 1.8 ± 1.2 days after the procedure, with no significant MR on pre-discharge transthoracic echocardiography. In 8 patients who underwent 1-month symptomatic assessment, clinical symptoms were diminished to New York Heart Association functional class I or II.

CONCLUSIONS

Transcatheter deployment of the ADO II plug was effective for the reduction of residual commissural MR and intraclip MR after MitraClip implantation. The potential role of this technique should be established for challenging cases.

Для цитирования: Kubo Shunsuke, Cox Justin M, Mizutani Yukiko, Uberoi Abhimanyu, Chakravarty Tarun, Nakajima Yoshifumi, Hussaini Asma, Tat Emily, Makar Moody, Kar Saibal. Transcatheter Procedure for Residual Mitral Regurgitation After MitraClip Implantation Using Amplatzer Duct Occluder II.. JACC. Cardiovascular interventions. 2016. DOI: 10.1016/j.jcin.2016.03.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.