MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jacc.2016.03.591

Медицинская библиотека · Journal of the American College of Cardiology · 2016

Treatment of Chronic Functional Mitral Valve Regurgitation With a Percutaneous Annuloplasty System.

Nickenig Georg, Schueler Robert, Dager Antonio, Martinez Clark Pedro, Abizaid Alexandre, Siminiak Tomasz, Buszman Pawel, Demkow Marcin, Ebner Adrian, Asch Federico M, Hammerstingl Christoph

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Current surgical and medical treatment options for functional mitral regurgitation (FMR) are limited and additional interventional approaches are required.

OBJECTIVES

This study sought to report the safety and performance data from the feasibility study with a novel direct annuloplasty system.

METHODS

Seventy-one patients with moderate to severe FMR (mean 67.7 ± 11.3 years of age, left ventricular [LV] ejection fraction 34.0 ± 8.3%), on stable medical heart failure medication were prospectively enrolled.

RESULTS

Device success rate was 70.4% (n = 50 of 71). No intraprocedural death occurred. In patients receiving implants, 4 patients (8.9%) experienced cardiac tamponade. Thirty-day (n = 45) and 6-month (n = 41) rates for all-cause mortality, stroke, and myocardial infarction were 4.4%, 4.4%, and 0.0% and 12.2%, 4.9%, and 0%, respectively. At 6 months, nonurgent mitral surgery was performed in 1 patient (2.4%) and nonurgent percutaneous repair in 7 patients (17.1%). Echocardiographic core analysis after 6 months showed mitral regurgitation reduction in 50% of treated patients by a mean of 1.3 grades. Concerning mitral valve (MV) annular geometry, we found significant reduction of anterior-posterior (-0.31 ± 0.4 cm) and septal-lateral dimensions (-0.21 ± 0.3 cm), a decreased MV-tenting area (-0.57 ± 1.1 cm(2)) and increase in MV coaptation length (0.13 ± 0.2 cm). Transthoracic echocardiography indicated reverse LV remodeling with reduction of LV end-diastolic diameter (-0.20 ± 0.4 mm) and volume (-22 ± 39 ml). Treatment was associated with significant improvement in 6-min walking distances (56.5 ± 92.0 m) and improvements in New York Heart Association functional class III/IV at 6 months from 53.3% to 23.3%.

CONCLUSIONS

Percutaneous direct annuloplasty is feasible and safe in high-risk FMR patients. This treatment initiates LV reverse remodeling, and provides clinical improvement during 6 months after treatment. (Mitralign Percutaneous Annuloplasty First in Man Study; NCT01852149).

Для цитирования: Nickenig Georg, Schueler Robert, Dager Antonio, Martinez Clark Pedro, Abizaid Alexandre, Siminiak Tomasz, Buszman Pawel, Demkow Marcin, Ebner Adrian, Asch Federico M, Hammerstingl Christoph. Treatment of Chronic Functional Mitral Valve Regurgitation With a Percutaneous Annuloplasty System.. Journal of the American College of Cardiology. 2016. DOI: 10.1016/j.jacc.2016.03.591.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.