Three-dimensional esophageal varix model quantification of variceal volume by high-resolution endoluminal US.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Chung C Y, McCray W H, Dhaliwal S, Haywood T, Black M, Liu J B, Miller L S

Аннотация

BACKGROUND: The purpose of this study was to evaluate the accuracy and reproducibility of three-dimensional volume measurements by high-resolution endoluminal ultrasound in an esophageal varix model.

METHODS: An esophageal varix model was made by filling three esophageal dilatation catheters with various volumes of water. A 20 MHz ultrasonography transducer was then pulled along the length of the catheters at a constant rate (1.25 mm/sec) while videotaping the procedure. Cross-sectional surface area measurements of each catheter were taken every second and the cross-sectional surface area was multiplied by the length of each catheter, as determined by high-resolution endoluminal ultrasound, to determine the volume in each catheter. Interobserver variability was calculated, and three-dimensional reconstruction was performed.

RESULTS: The measured volumes corresponded closely with the actual volumes with an error ranging from 0% to 15.4%. The correlation between actual and measured volumes was r = 0.988. The interobserver variability ranged from r = 0.951 to r = 0.994. Actual esophageal varices were then imaged in a similar fashion to determine the feasibility of this method in patients with esophageal varices.

CONCLUSIONS: High-resolution endoluminal ultrasound is an accurate and reproducible method of measuring volumes in an esophageal varix model and can be used in a clinical setting to determine variceal volume. Volume studies are now underway in human subjects.