Eichinger Walter B, Botzenhardt Florian, Guenzinger Ralf, Bleiziffer Sabine, Keithahn Alexandra, Bauernschmitt Robert, Lange Ruediger

Клиническое исследованиеСравнительное исследованиеСтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIM OF THE STUDY: The aim of this prospective, randomized study was to compare the hemodynamic performance of the Medtronic Mosaic and Edwards Perimount bioprostheses in the aortic position, and to evaluate prosthesis-specific differences in valve sizing and valve-size labeling.

METHODS

Between August 2000 and September 2002, 139 patients underwent isolated aortic valve replacement (AVR) with the Mosaic (n = 67) or Perimount (n = 72) bioprosthesis. Intraoperatively, the internal aortic annulus diameter was measured by insertion of a gauge (Hegar dilator), while prosthesis size was determined by using the original sizers. Transthoracic echocardiography was performed to determine hemodynamic and dimensional data. As the aim of AVR is to achieve a maximal effective orifice area (EOA) within a given aortic annulus, the ratio of EOA to patient aortic annulus area was calculated, the latter being based on annulus diameter measured intraoperatively.

RESULTS

Operative mortality was 2.2% (Mosaic 3.0%; Perimount 1.4%; p = NS). Upsizing (using a prosthesis larger in labeled valve size than the patient's measured internal aortic annulus diameter) was possible in 28.4% of Mosaic patients and 8.3% of Perimount patients. The postoperative mean systolic pressure gradient ranged from 10.5 to 22.2 mmHg in the Mosaic group, and from 9.4 to 12.6 mmHg in the Perimount group; it was significantly lower for 21 and 23 Perimount valves than for 21 and 23 Mosaic valves. The EOA ranged from 0.78 to 2.37 cm2 in Mosaic patients, and from 0.95 to 2.12 cm2 in Perimount patients. When indexing EOA by calculating the ratio of EOA to patient aortic annulus area to adjust for variables such as patient anatomy and valve dimensions, there was no significant difference between the two bioprostheses.

CONCLUSION

Comparisons of absolute EOA values grouped by the manufacturers' valve sizes are misleading because of specific differences in geometric dimensions. The EOA:patient aortic annulus area ratio provides a new hemodynamic index which may facilitate objective comparisons between different valve types.

Для цитирования: Eichinger Walter B, Botzenhardt Florian, Guenzinger Ralf, Bleiziffer Sabine, Keithahn Alexandra, Bauernschmitt Robert, Lange Ruediger. The effective orifice area/patient aortic annulus area ratio: a better way to compare different bioprostheses? A prospective randomized comparison of the Mosaic and Perimount bioprostheses in the aortic position.. The Journal of heart valve disease. 2004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.