MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/heart.89.11.1316

Медицинская библиотека · Heart (British Cardiac Society) · 2003

Prevalence and clinical significance of incidental paraprosthetic valvar regurgitation: a prospective study using transoesophageal echocardiography.

Ionescu A, Fraser A G, Butchart E G

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To assess the prevalence, mechanisms, and significance of paraprosthetic regurgitation detected incidentally by transoesophageal echocardiography (TOE) in patients after heart valve replacement.

DESIGN

Prospective observational study.

SETTING

Tertiary referral centre.

PATIENTS: 360 consecutive patients (mean (SD) age 65.8(9.5) years, 193 women) undergoing elective first ever valve replacement.

METHODS

Postoperative and follow up TOE, and tests for haemolysis and anaemia.

RESULTS

There were 243 aortic, 90 mitral, and 27 double valve replacements, using 316 mechanical and 44 tissue valves, giving 270 aortic and 117 mitral valves. One patient with severe paraprosthetic mitral regurgitation underwent immediate reoperation and was excluded from subsequent analyses. Paraprosthetic jets were detected around 16 (6%) of the aortic and 38 (32%) of the mitral valves (p < 0.05) at the postoperative study. Follow up TOE was available for 151 aortic and 67 mitral valves, 0.9 (0.5) years after operation. Paraprosthetic jets were present in 15 (10%) of the aortic and 10 (15%) of the mitral valves (NS). Two thirds of the aortic and a fifth of the mitral jets were new. Paraprosthetic jets were more common in aortic valves in a supra-annular (12 of 88, 14%) than in an intra-annular position (4 or 182, 2%; p < 0.005) and in mitral valves inserted with continuous (36 of 88, 41%) rather than interrupted sutures (2 of 28, 7%; p < 0.001). Lactate dehydrogenase concentration was higher in patients with paraprosthetic jets than in those without (752 (236) v 654 (208) IU/l, p < 0.001). Haemoglobin and haptoglobin concentrations were not different.

CONCLUSIONS

Small paraprosthetic leaks are common, are related to surgical factors, are not associated with increased subclinical haemolysis, and are benign during the first year after heart valve replacement.

Для цитирования: Ionescu A, Fraser A G, Butchart E G. Prevalence and clinical significance of incidental paraprosthetic valvar regurgitation: a prospective study using transoesophageal echocardiography.. Heart (British Cardiac Society). 2003. DOI: 10.1136/heart.89.11.1316.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.