DOI 10.1161/circep.119.007377

Медицинская библиотека · Circulation. Arrhythmia and electrophysiology · 2020

Evaluation of ECG Imaging to Map Hemodynamically Stable and Unstable Ventricular Arrhythmias.

Graham Adam J, Orini Michele, Zacur Ernesto, Dhillon Gurpreet, Daw Holly, Srinivasan Neil T, Martin Claire, Lane Jem, Mansell Josephine S, Cambridge Alex, Garcia Jason, Pugliese Francesca, Segal Oliver, Ahsan Syed, Lowe Martin, Finlay Malcolm, Earley Mark J, Chow Anthony, Sporton Simon, Dhinoja Mehul, Hunter Ross J, Schilling Richard J, Lambiase Pier D

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

ECG imaging (ECGI) has been used to guide treatment of ventricular ectopy and arrhythmias. However, the accuracy of ECGI in localizing the origin of arrhythmias during catheter ablation of ventricular tachycardia (VT) in structurally abnormal hearts remains to be fully validated.

METHODS

During catheter ablation of VT, simultaneous mapping was performed using electroanatomical mapping (CARTO, Biosense-Webster) and ECGI (CardioInsight, Medtronic) in 18 patients. Sites of entrainment, pace-mapping, and termination during ablation were used to define the VT site of origin (SoO). Distance between SoO and the site of earliest activation on ECGI were measured using co-registered geometries from both systems. The accuracy of ECGI versus a 12-lead surface ECG algorithm was compared.

RESULTS

A total of 29 VTs were available for comparison. Distance between SoO and sites of earliest activation in ECGI was 22.6, 13.9 to 36.2 mm (median, first to third quartile). ECGI mapped VT sites of origin onto the correct AHA segment with higher accuracy than a validated 12-lead ECG algorithm (83.3% versus 38.9%; P=0.015).

CONCLUSIONS

This simultaneous assessment demonstrates that CardioInsight localizes VT circuits with sufficient accuracy to provide a region of interest for targeting mapping for ablation. Resolution is not sufficient to guide discrete radiofrequency lesion delivery via catheter ablation without concomitant use of an electroanatomical mapping system but may be sufficient for segmental ablation with radiotherapy.

Для цитирования: Graham Adam J, Orini Michele, Zacur Ernesto, Dhillon Gurpreet, Daw Holly, Srinivasan Neil T, Martin Claire, Lane Jem, Mansell Josephine S, Cambridge Alex, Garcia Jason, Pugliese Francesca, Segal Oliver, Ahsan Syed, Lowe Martin, Finlay Malcolm, Earley Mark J, Chow Anthony, Sporton Simon, Dhinoja Mehul, Hunter Ross J, Schilling Richard J, Lambiase Pier D. Evaluation of ECG Imaging to Map Hemodynamically Stable and Unstable Ventricular Arrhythmias.. Circulation. Arrhythmia and electrophysiology. 2020. DOI: 10.1161/circep.119.007377.