m+Medelevator.RU
← Поиск публикаций
medRxiv : the preprint server for health sciences · 2025

Obese patient imaging: improved dose efficiency with photon-counting CT.

СтатьяПрепринтНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Liu Leening P, Mei Kai, Sharma Shobhit, Ross Steven, Halliburton Sandra S, Thompson Richard, Akino Naruomi, Dhanaliwala Ali H, Roshkovan Leonid, Litt Harold I, Noël Peter B

Аннотация

OBJECTIVE: To evaluate the dose efficiency of cadmium-zinc-telluride (CZT) based photon-counting CT (PCCT) compared to energy-integrating detector CT (EID-CT) across phantom sizes.

METHODS: A patient-specific 3D-printed pancreas phantom and a phantom with tissue mimicking inserts were placed in extension rings corresponding to the 50th, 75th, 85th, and 95th percentile adult waist circumferences. Phantoms were scanned on both PCCT and EID-CT with CTDIvol ranging from 0.5 to 19.4 mGy. Noise was measured in both phantoms to evaluate dose efficiency. Non-Poisson noise at low doses (<2 mGy) was quantified using root mean square error from linear fits of the noise-dose relationship. Potential dose reduction was then assessed by matching noise levels between scanners across phantom sizes.

RESULTS: PCCT demonstrated reduced noise compared to EID-CT across all phantom sizes and doses with average noise reductions of 22%, 23%, 25%, and 28% for the 50th, 75th, 85th, and 95th percentile phantoms, respectively. Noise reduction intensified at lower doses and larger phantom sizes, reaching 88 HU at 1 mGy for the 95th percentile phantom. Non-Poisson noise decreased significantly with PCCT compared to EID-CT for all phantom sizes (p < 0.013). At matched noise levels, PCCT enabled dose reductions of 33% and 44% for the 50th and 95th percentile phantoms, respectively.

CONCLUSIONS: PCCT exhibited superior dose efficiency compared to EID-CT across a range of phantom sizes. The enhanced dose efficiency enables both noise reduction and potential dose reduction for the imaging of obese patients and low-dose imaging applications.

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты