MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.25862

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2015

First-in-man study evaluating the safety and efficacy of a second generation biodegradable polymer sirolimus-eluting stent in the treatment of patients with de novo coronary lesions: clinical, Angiographic, and OCT outcomes of CREDIT-1.

Wang Geng, Sun Zhongwei, Jin Quanmin, Xu Kai, Li Yi, Wang Xiaozeng, Ma Yingyan, Liu Haiwei, Zhao Xin, Wang Bin, Deng Jie, Guan Shaoyi, Ge Meiling, Wang Xiaoyan, Xu Bo, Han Yaling

СтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the preliminary safety and efficacy of the EXCEL II stent system.

BACKGROUND

Although the first biodegradable polymer drug-eluting stent (BP-DES), EXCEL, was launched nearly a decade ago, in-stent restenosis and stent thrombosis remain pertinent clinical problems in practice. A new cobalt-chromium BP-DES EXCEL II has been developed with the aim of improving stent safety and efficacy.

METHODS

Forty-five patients with single de novo native coronary lesions were enrolled and randomized to two groups in a 2:1 ratio, the 4-month follow-up group (n = 30) and the 12-month follow-up group (n = 15). All patients underwent percutaneous coronary intervention (PCI) with the EXCEL II stent system. Quantitative coronary angiography (QCA) and optical coherence tomography (OCT) were used to assess coronary vasculature at the designated 4- or 12-month follow-up. The primary outcome was major adverse cardiac events (MACE) at 30 days post-PCI.

RESULTS

No MACE, thrombotic events, or target lesion failure was found in the 45 patients during the 12-month follow-up. There was no significant difference (P > 0.05) between the two groups in terms of in-stent and in-segment late lumen loss (LLL). No in-stent and in-segment restenosis was found in either group. At follow-up, the ratio of >10% uncovered struts per lesion was 26.67% in the 4-month group and 0% in the 12-month group (P < 0.05). Neointimal coverage in the 12-month group was significantly better than in the 4-month group (98.58% vs. 93.51%, P < 0.01).

CONCLUSIONS

This first-in-man study demonstrates promising feasibility, safety, and efficacy of EXCEL II stents. These stents were found to have rapid endothelialization and low LLL rates at 4 and 12 months after implantation.

Для цитирования: Wang Geng, Sun Zhongwei, Jin Quanmin, Xu Kai, Li Yi, Wang Xiaozeng, Ma Yingyan, Liu Haiwei, Zhao Xin, Wang Bin, Deng Jie, Guan Shaoyi, Ge Meiling, Wang Xiaoyan, Xu Bo, Han Yaling. First-in-man study evaluating the safety and efficacy of a second generation biodegradable polymer sirolimus-eluting stent in the treatment of patients with de novo coronary lesions: clinical, Angiographic, and OCT outcomes of CREDIT-1.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2015. DOI: 10.1002/ccd.25862.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.