DOI 10.1016/j.jaccas.2026.110194

Медицинская библиотека · JACC. Case reports · 2026

Subacute Apical RV Lead Microperforation in CRT Implantation: An Electrical Diagnosis.

Halder Ashesh, Dutta Suvradip, Kulshrestha Srashti, Chaudhuri Arya, Pandey Anand Kumar

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Right ventricular (RV) lead perforation is a rare complication of cardiac implantable electronic device implantation and may be difficult to diagnose when imaging findings are absent or subtle.

CASE SUMMARY: A 54-year-old man developed positional diaphragmatic stimulation 3 days after cardiac resynchronization therapy implantation. Device interrogation demonstrated rising RV pacing thresholds, declining lead impedance, and marked disparity between unipolar and bipolar parameters, with loss of unipolar tip capture despite preserved bipolar capture. Transthoracic echocardiography and fluoroscopy showed no lead displacement or pericardial effusion. Based on these electrical findings, RV lead microperforation was suspected. Lead revision resulted in normalization of device parameters and complete symptom resolution.

DISCUSSION: Microperforation may present without radiographic abnormalities, making diagnosis challenging. This case illustrates how configuration-dependent electrical abnormalities can distinguish microperforation from other causes of lead dysfunction and facilitate early intervention.

TAKE-HOME MESSAGE: Careful interpretation of cardiac implantable electronic device electrical parameters, together with clinical assessment, can enable early diagnosis of lead microperforation even when conventional imaging is unrevealing.

Для цитирования: Halder Ashesh, Dutta Suvradip, Kulshrestha Srashti, Chaudhuri Arya, Pandey Anand Kumar. Subacute Apical RV Lead Microperforation in CRT Implantation: An Electrical Diagnosis.. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.110194.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.