Akita Masafumi, Takenaka Hiroaki, Harada Masakazu, Li Tao-Sheng, Zhang Hua, Zempo Nobuya, Esato Kensuke, Nagano Tomokazu, Taiji Mutsuo, Noguchi Hiroshi
Аннотация
PURPOSE: To evaluate whether nonpenetrating vascular closure staples (VCS) and hepatocyte growth factor (HGF) can effectively prevent anastomotic intimal hyperplasia.
METHODS: An expanded polytetrafluoroethylene graft, 2 mm in diameter, was implanted in the common carotid artery of rabbits divided into three experimental groups. In the control group, distal anastomosis was performed with interrupted suturing; in the VCS group, clips were applied along the lateral suture line after the placement of stay sutures; and in the VCS + HGF group, the same anastomotic technique was performed as in the VCS group, followed by the administration of the HGF for 4 days.
RESULTS: The time taken to complete the anastomosis was significantly less in both the VCS groups than in the control group ( P < 0.0001). On postoperative day (POD) 28, the patency rate was significantly lower ( P < 0.05) in the VCS group (42.9%) than in the control group (100%), but the rate in the VCS + HGF group (100%) was the same as that in the control group. Intimal thickness was significantly less in the control group than in either the VCS or VCS + HGF groups ( P< 0.05). The percentage of area stenosis was significantly less ( P< 0.01) in the control group than in the VCS group.
CONCLUSION: The VCS clip failed to suppress intimal thickness or reduce the percentage of stenosis at the anastomotic site.
METHODS: An expanded polytetrafluoroethylene graft, 2 mm in diameter, was implanted in the common carotid artery of rabbits divided into three experimental groups. In the control group, distal anastomosis was performed with interrupted suturing; in the VCS group, clips were applied along the lateral suture line after the placement of stay sutures; and in the VCS + HGF group, the same anastomotic technique was performed as in the VCS group, followed by the administration of the HGF for 4 days.
RESULTS: The time taken to complete the anastomosis was significantly less in both the VCS groups than in the control group ( P < 0.0001). On postoperative day (POD) 28, the patency rate was significantly lower ( P < 0.05) in the VCS group (42.9%) than in the control group (100%), but the rate in the VCS + HGF group (100%) was the same as that in the control group. Intimal thickness was significantly less in the control group than in either the VCS or VCS + HGF groups ( P< 0.05). The percentage of area stenosis was significantly less ( P< 0.01) in the control group than in the VCS group.
CONCLUSION: The VCS clip failed to suppress intimal thickness or reduce the percentage of stenosis at the anastomotic site.