DOI 10.1016/j.clinph.2005.04.006
Effect of amplifier gain setting on distal motor latency in normal subjects and CTS patients.
Аннотация
OBJECTIVE
To determine normative cutoffs and sensitivities for median distal latency (MDL), median-thenar to ulnar-thenar latency difference (TTLD), and median-thenar to ulnar-hypothenar latency difference (THLD) at various amplifier gains for carpal tunnel syndrome (CTS) electrodiagnosis. A prior study utilized only an amplifier gain of 0.2 mV/division.
METHODS
Abnormal cutoffs for MDL, TTLD and THLD were determined based on 34 control hands at gains of 0.2, 0.5, 1, 2, and 5 mV. Diagnostic sensitivities were determined for 50 patients (80 hands) with clinically and electrodiagnostically defined CTS.
RESULTS
At a gain of 0.2 and 0.5 mV/division, abnormal cutoffs for MDL, THLD, and TTLD were: 3.7, 1.2, and 0.8 ms. At gains of 1, 2, and 5 mV the abnormal cutoffs were 4, 1.2, and 1 ms. The sensitivities at gains of 0.2, 0.5, 1, 2, and 5 mV for MDL, THLD, and TTLD were: 65, 66, 53, 57, 61/86, 83, 88, 86, 86/91, 91, 76, 73, 59.
CONCLUSIONS
MDL and THLD sensitivities are gain-independent. THLD is substantially more sensitive than MDL at all gains. TTLD sensitivity is maximized with 0.2 and 0.5 mV gains.
SIGNIFICANCE: TTLD and THLD increase diagnostic sensitivity with minimal additional effort. TTLD sensitivity is maximized with 0.2 or 0.5 mV gains. The electromyographer's preferred gain may be used.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.