Initial Experience Using the Gore Embolic Filter in Carotid Interventions.
Аннотация
BACKGROUND
This is the first clinical report on experience in the use of the Gore embolic filter in carotid interventions. It was designed as a guidewire and embolic protection system in carotid, peripheral, and coronary interventions. The ability to capture debris is driven by the frame of the filter, which is designed to improve vessel wall apposition and allows a short landing zone.
METHODS
We report the results of the first 20 consecutive patients undergoing carotid artery stenting using the Gore embolic filter in our institution. We analyzed technical success as well as the occurrence of transient ischemic attack (TIA), stroke, or death periprocedurally and through 30 days of follow-up. Mean patient age was 72 years and 12 patients (60%) were male. Seven patients were symptomatic and 4 patients suffered recurrent neurological events.
RESULTS
Technical success was achieved in all procedures. In 1 patient, the retrieval catheter was caught between the proximal struts of the stent and required further retrieval maneuvers. Within 30 days of follow-up, 1 patient had a TIA. No stroke, death, or myocardial infarction occurred.
CONCLUSION
This initial experience suggests that the Gore embolic filter device can be used safely for distal embolic protection during carotid stenting procedures with high technical success.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.