DOI 10.1016/j.cca.2010.01.034

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2010

Comparison of amino acids and acylcarnitines assay methods used in newborn screening assays by tandem mass spectrometry.

De Jesús Víctor R, Chace Donald H, Lim Timothy H, Mei Joanne V, Hannon W Harry

Сравнительное исследованиеСтатьяПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The analysis of amino acids (AA) and acylcarnitines (AC) by tandem mass spectrometry (MS/MS) is performed in newborn screening laboratories worldwide. While butyl esterification assays are routine, it is possible to detect AAs and ACs as their native free acids (underivatized). The Centers for Disease Control and Prevention's Newborn Screening Quality Assurance Program provides dried blood spot (DBS) quality control (QC) and proficiency testing (PT) programs for numerous MS/MS analytes. We describe empirical differences between derivatization and non-derivatization techniques for selected AAs and ACs.

METHODS

DBS materials were prepared at levels near, above and below mean domestic laboratory cut-offs, and distributed to program participants for MS/MS analysis. Laboratories reported quantitative and qualitative results. QC DBS materials were assayed in-house following established protocols.

RESULT: Minor differences (<15%) between quantitative values resulting from butyl esters and free acid techniques were observed for the majority of the analytes. Mass spectrometric response from underivatized dicarboxylic acid acylcarnitines was less intense than their butyl esters.

CONCLUSIONS

The use of underivatized techniques may also result in the inability to differentiate isobaric acylcarnitines. Laboratories should establish their own protocols by focusing on the decisions that identify test results requiring additional follow-up testing versus those that do not.

Для цитирования: De Jesús Víctor R, Chace Donald H, Lim Timothy H, Mei Joanne V, Hannon W Harry. Comparison of amino acids and acylcarnitines assay methods used in newborn screening assays by tandem mass spectrometry.. Clinica chimica acta; international journal of clinical chemistry. 2010. DOI: 10.1016/j.cca.2010.01.034.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.