MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jaccas.2026.110182

Медицинская библиотека · JACC. Case reports · 2026

Loss of Biventricular Capture After CRT-D Implantation: Device Malfunction or Something Else?

Weber Paul, Govil Dhruva, Shah Dipak

СтатьяПеревод в очереди

Аннотация

CASE SUMMARY: A 75-year-old woman with nonischemic cardiomyopathy and left bundle branch block underwent cardiac resynchronization therapy defibrillator implantation. Postimplant electrocardiography demonstrated successful biventricular pacing with QRS narrowing to 110 ms from a preimplant baseline of 169 ms. Three consecutive beats with an abruptly different QRS morphology raised concern for intermittent loss of biventricular pacing or device malfunction.

TAKE-HOME MESSAGE: Dynamic atrioventricular-delay optimization algorithms such as SyncAV can produce transient changes in QRS morphology that may mimic loss of biventricular pacing but represent expected device behavior and recognition of this pattern prevents unnecessary device evaluation.

Для цитирования: Weber Paul, Govil Dhruva, Shah Dipak. Loss of Biventricular Capture After CRT-D Implantation: Device Malfunction or Something Else?. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.110182.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.