Incidence and predictors of persistent left atrial appendage patency and its subtypes after percutaneous closure: a pre-specified analysis of the SWISS-APERO trial.

Galea Roberto, Perich Krsnik Juan, Peters Alan Arthur, De Marco Federico, Aminian Adel, Meneveau Nicolas, Heg Dierik, Gräni Christoph, Anselme Frederic, Franzone Anna, Vranckx Pascal, Fischer Urs, Bedogni Francesco, Räber Lorenz, Valgimigli Marco

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIMS: Left atrial appendage (LAA) patency after percutaneous closure is associated with higher thromboembolic risk, especially when detected at later time points after procedure. We aim to investigate the incidence and predictors of persistent LAA patency and its different subtypes at cardiac computed tomography angiography (CCTA) during the first year after LAA closure (LAAC).

METHODS AND

RESULTS

In the SWISS APERO trial, patients undergoing LAAC were randomly assigned (1:1) to Amulet or Watchman/FLX across eight European centers and received CCTA follow-up at 45 days and 13 months. The presence, the type and predictors of persistent or new-onset LAA patency at 13 months were investigated by matching 45-day and 13-month CCTAs in an imaging core-lab. Of the 221 randomized patients, 158 (71.5%) were included in the matched 45-day and 13-month CCTA analysis. Residual LAA patency decreased from 66.4 to 50.6% with approximately one-third (30.5%) of 45-day patent appendages (PA) resolved at 13 months. Side-gap leaks at 45 days were the PA subtypes associated with the highest likelihood (86%) of persistence. At 45-day CCTA, the presence of at least two factors among side-gap leak proximal area >37.7 mm2, radial diameter >3 mm or LAA-device angle >39.4° ('ARA LAAO' score ≥ 2) predicted leak persistence at 13-month CCTA with 94.1% sensitivity and 70% specificity.

CONCLUSION

In a prospective multi-centre cohort of clinically indicated LAAC, approximately two-thirds of 45-day PAs, especially those related to large side-gap leaks with bigger LAA-device off-axis angle, persisted at 13 months, whereas new-onset PA occurred in roughly 1 in every 10 patients. Clinical Trial Registration: URL https://clinicaltrials.gov Unique Identifier NCT03399851.

Для цитирования: Galea Roberto, Perich Krsnik Juan, Peters Alan Arthur, De Marco Federico, Aminian Adel, Meneveau Nicolas, Heg Dierik, Gräni Christoph, Anselme Frederic, Franzone Anna, Vranckx Pascal, Fischer Urs, Bedogni Francesco, Räber Lorenz, Valgimigli Marco. Incidence and predictors of persistent left atrial appendage patency and its subtypes after percutaneous closure: a pre-specified analysis of the SWISS-APERO trial.. European heart journal. Cardiovascular Imaging. 2025. DOI: 10.1093/ehjci/jeaf232.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.