m+Medelevator.RU
← Поиск публикаций
Magnetic resonance in medicine · 2023

B1 and magnetization decay correction for hyperpolarized 129 Xe lung imaging using sequential 2D spiral acquisitions.

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Bdaiwi Abdullah S, Costa Mariah L, Plummer Joseph W, Willmering Matthew M, Walkup Laura L, Cleveland Zackary I

Аннотация

PURPOSE: To mitigate signal variations caused by inhomogeneous RF and magnetization decay in hyperpolarized 129 Xe ventilation images using flip-angle maps generated from sequential 2D spiral ventilation images acquired in a breath-hold. Images and correction maps were compared with those obtained using conventional, 2D gradient-recalled echo.

THEORY AND METHODS: Analytical expressions to predict signal intensity and uncertainty in flip-angle measurements were derived from the Bloch equations and validated by simulations and phantom experiments. Imaging in 129 Xe phantoms and human subjects (1 healthy, 1 cystic fibrosis) was performed using 2D gradient-recalled echo and spiral. For both sequences, consecutive images were acquired with the same slice position during a breath-hold (Cartesian scan time = 15 s; spiral scan time = 5 s). The ratio of these images was used to calculate flip-angle maps and correct intensity inhomogeneities in ventilation images.

RESULTS: Mean measured flip angle showed excellent agreement with the applied flip angle in simulations (R2  = 0.99) for both sequences. Mean measured flip angle agreed well with the globally applied flip angle (∼15% difference) in 129 Xe phantoms and in vivo imaging using both sequences. Corrected images displayed reduced coil-dependent signal nonuniformity relative to uncorrected images.

CONCLUSIONS: Flip-angle maps were obtained using sequentially acquired, 2D spiral, 129 Xe ventilation images. Signal intensity variations caused by RF-coil inhomogeneity can be corrected by acquiring sequential single-breath ventilation images in less than 5-s scan time. Thus, this method can be used to remove undesirable heterogeneity while preserving physiological effects on the signal distribution.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты