DOI 10.1177/23259671211021820

Медицинская библиотека · Orthopaedic journal of sports medicine · 2021

Suture Slippage After Arthroscopic Cuff Repair: Medial Displacement of Suture Knots on Follow-up Ultrasonography.

Kim Hyungsuk, Han Sung Bin, Song Hyun Seok

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Compared with the single-row technique, the double-row rotator cuff repair technique is known to have a higher load to failure and a lower frequency of gap formation, leading to a lower retear rate. There are some patients with poor clinical outcomes or poor muscle strength without radiologic retear.

PURPOSE/HYPOTHESIS: To assess the postoperative position of suture knots via serial ultrasonography in patients who had undergone arthroscopic rotator cuff repair with the suture-bridge technique. Our hypothesis was that the suture would pull out of the lateral anchor (suture slippage), changing the positions of the medial suture knots during healing.

STUDY

DESIGN

Case series; Level of evidence, 4.

METHODS

This study included 53 patients (55 shoulders) who underwent arthroscopic suture-bridge repair and were evaluated for a minimum of 24 months. On serial ultrasonography, a straight line was drawn between the top of the greater tuberosity and the medial cortex of the anchor hole. The distances between the knots of the medial rows and the perpendicular line through the center of the anchor hole were measured in longitudinal plane images of the supraspinatus. Follow-up ultrasonography was performed at 2, 3, and 6 months postoperatively as well as at the final visit. The visual analog scale, the American Shoulder and Elbow Surgeons score, the Constant score, and the University of California, Los Angeles shoulder score were recorded preoperatively and on the final follow-up.

RESULTS

Of the 55 shoulders, 6 developed retears at repaired sites. The mean follow-up duration was 37.5 months (range, 24-65 months). Slippage distance increased significantly over time (P < .001). The slippage at the final visit did not differ between patients with retear and no retear (13.4 mm for retear group; 10.6 mm for no retear group [P = .096]).

CONCLUSION

Suture knots of the medial row migrated medially via a suture pullout from the lateral row anchor of suture-bridge technique. Suture slippage distance did not differ significantly between retear and no retear groups.

Для цитирования: Kim Hyungsuk, Han Sung Bin, Song Hyun Seok. Suture Slippage After Arthroscopic Cuff Repair: Medial Displacement of Suture Knots on Follow-up Ultrasonography.. Orthopaedic journal of sports medicine. 2021. DOI: 10.1177/23259671211021820.