DOI 10.1016/j.ijcard.2021.07.050
Morphologic investigation on Perceval S, a sutureless pericardial valve prosthesis: collagen integrity after collapsing-ballooning and structural valve deterioration at distance.
Аннотация
Perceval S is a self-expandable, stent-mounted bioprosthetic valve (BPV), with glutaraldehyde treated bovine pericardium, processed with homocysteic acid as an anti-calcification treatment. The stent is crimpable but the valve insertion is done surgically via a shorter procedure which does not require sutures.
OBJECTIVES
MATERIAL AND
METHODS
RESULTS
CONCLUSIONS
Collapsing and ballooning do not alter cusp collagen periodicity. Structural valve deterioration with stenosis, due to dystrophic calcification and fibrous tissue overgrowth, seldom occurred in the mid-term. Glutaraldehyde fixed pericardium has the potential to undergo structural valve deterioration with time, similar to well-known BPV failure. This supports the recommendation to pursue improvement of tissue valve treatment with enhanced durability.