Evaluation of an ambulatory device, CID 102, in the diagnosis of obstructive sleep apnoea syndrome.

Van Surell C, Lemaigre D, Leroy M, Foucher A, Hagenmuller M P, Raffestin B

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Diagnosis of obstructive sleep apnoea syndrome (OSAS) is usually performed during overnight polysomnography in the sleep laboratory. In an attempt to simplify the diagnostic strategy, we compared an ambulatory device, CID 102, with polysomnography during the same night in the laboratory in 50 consecutive patients referred for polysomnography. The CID 102 device monitors oxygen saturation, heart rate, body position and tracheal breath sounds. An acoustic pressure sensor is placed on the suprasternal notch. Signals coming from this sensor are amplified and analysed in three different channels, according to their frequency and energy. CID respiratory disturbance index is defined as the number, per hour of analysis time, of apnoeas lasting more than 10 s plus episodes of desaturation by 4% or more associated with pauses lasting from 7-10 s or snores. The polysomnographic data were recorded on paper (Reega 2000, Alvar) and analysed manually. Polysomnographic apnoea-hypopnoea index (AHIp) was defined as the number of apnoeas plus hypopnoeas per hour of sleep. The sensitivity, specificity, positive predictive value and negative predictive value of various CID respiratory disturbance index (> or = 5, > or = 10, > or = 15 and > or = 20 per hour) in diagnosing obstructive sleep apnoea syndrome were determined. When OSAS was diagnosed as AHIp > or = 15, sensitivity and specificity of a CID respiratory disturbance index > or = 5 were 73 and 62%, respectively. Positive predictive value of CID respiratory disturbance index > or = 10 for AHIp > or = 10 was 94%. CID 102 false negative patients had only hypopnoeas without any desaturation.(ABSTRACT TRUNCATED AT 250 WORDS)

Для цитирования: Van Surell C, Lemaigre D, Leroy M, Foucher A, Hagenmuller M P, Raffestin B. Evaluation of an ambulatory device, CID 102, in the diagnosis of obstructive sleep apnoea syndrome.. The European respiratory journal. 1995.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.