Acute procedural and cryoballoon characteristics from cryoablation of atrial fibrillation using the first- and second-generation cryoballoon: a retrospective comparative study with follow-up outcomes.
Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Aryana Arash, Morkoch Shemsa, Bailey Sean, Lim Hae W, Sara Rahmani, d'Avila André, O'Neill P Gearoid
Аннотация
PURPOSE: There is limited data available on the safety and efficacy of the second-generation cryoballoon (CB-2) for cryoablation of atrial fibrillation (Cryo-AF). We evaluated the procedural, biophysical, and clinical outcomes of Cryo-AF in a large patient cohort using CB-2 as compared with the first-generation cryoballoon (CB-1).
METHODS: Three-hundred and forty consecutive patients undergoing Cryo-AF with CB-1 (n = 140) and CB-2 (n = 200) were retrospectively evaluated.
RESULTS: Paroxysmal AF was more prevalent in CB-1 (86%) versus CB-2 (72%) (p = 0.001). During Cryo-AF, the mean balloon temperature was lower with CB-2 at 30 s (8 versus -4°C; p < 0.001) and 60 s (-26 versus -32°C; p < 0.001) with equivalent nadir temperatures (both at -50°C; p = 0.542). With CB-2, time-to-nadir temperature was shorter (232 versus 209 s; p < 0.001) and thaw times were longer (47 versus 53 s; p < 0.001). Acute pulmonary vein (PV) isolation rate was higher with CB-2 (92 versus 98%; p = 0.036) despite reduced cryoablation time (61 versus 47 min; p < 0.001) and freeze area-under-the-curve (-155,044 versus -116,740 s°C; p < 0.001). With CB-2, procedure time (209 versus 154 min; p < 0.001) and fluoroscopy time (42 versus 27 min; p < 0.001) were shorter, with similar acute/long-term adverse events (AEs) and freedom from AF at 6, 9, and 12 months (89, 86, and 82%) during 16 ± 8 months of follow-up. However, CB-2 was associated with lower PV reconnection rates at redo ablation (30 versus 13%; p = 0.037).
CONCLUSIONS: With CB-2, acute and long-term PV isolation rates were higher despite shorter ablations, faster balloon cooling, and longer thaw times, with similar AE rates and freedom from AF.
METHODS: Three-hundred and forty consecutive patients undergoing Cryo-AF with CB-1 (n = 140) and CB-2 (n = 200) were retrospectively evaluated.
RESULTS: Paroxysmal AF was more prevalent in CB-1 (86%) versus CB-2 (72%) (p = 0.001). During Cryo-AF, the mean balloon temperature was lower with CB-2 at 30 s (8 versus -4°C; p < 0.001) and 60 s (-26 versus -32°C; p < 0.001) with equivalent nadir temperatures (both at -50°C; p = 0.542). With CB-2, time-to-nadir temperature was shorter (232 versus 209 s; p < 0.001) and thaw times were longer (47 versus 53 s; p < 0.001). Acute pulmonary vein (PV) isolation rate was higher with CB-2 (92 versus 98%; p = 0.036) despite reduced cryoablation time (61 versus 47 min; p < 0.001) and freeze area-under-the-curve (-155,044 versus -116,740 s°C; p < 0.001). With CB-2, procedure time (209 versus 154 min; p < 0.001) and fluoroscopy time (42 versus 27 min; p < 0.001) were shorter, with similar acute/long-term adverse events (AEs) and freedom from AF at 6, 9, and 12 months (89, 86, and 82%) during 16 ± 8 months of follow-up. However, CB-2 was associated with lower PV reconnection rates at redo ablation (30 versus 13%; p = 0.037).
CONCLUSIONS: With CB-2, acute and long-term PV isolation rates were higher despite shorter ablations, faster balloon cooling, and longer thaw times, with similar AE rates and freedom from AF.