DOI 10.1016/j.hrthm.2014.05.014

Медицинская библиотека · Heart rhythm · 2014

Multicenter experience with extraction of the Riata/Riata ST ICD lead.

Maytin Melanie, Wilkoff Bruce L, Brunner Michael, Cronin Edmond, Love Charles J, Grazia Bongiorni Maria, Segreti Luca, Carrillo Roger G, Garisto Juan D, Kutalek Steven, Subzposh Faiz, Fischer Avi, Coffey James O, Gangireddy Sandeep R, Saba Samir, Mittal Suneet, Arshad Aysha, O'Keefe Ryan Michael, Henrikson Charles A, Belott Peter, John Roy M, Epstein Laurence M

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

In November 2011, the Food and Drug Administration issued a class I recall of Riata and Riata ST implantable cardioverter-defibrillator leads. Management recommendations regarding the recall have remained controversial.

OBJECTIVE

Data regarding the safety and feasibility of extraction of Riata implantable cardioverter-defibrillator leads are limited.

METHODS

We performed a retrospective study of patients undergoing extraction of Riata/Riata ST leads at 11 centers.

RESULTS

Between July 2003 and April 2013, 577 Riata/Riata ST leads were extracted from 577 patients (Riata 467, [84%]; Riata ST 89, [16%]). Complete procedural success achieved in 99.1%. The cohort was 78% men, with a mean age of 60 years and a mean left ventricular ejection fraction of 34% ± 14%. The mean implant duration was 44.7 months (range 0-124.6 months). The majority of leads extracted were for infection (305 [53.0%]) and 220 (35.7%) for lead malfunction. Evaluation for lead integrity was performed in 295 cases. Of these, 34.9% were found to have externalized cables. Implant duration was significantly longer in leads with externalized cables (P < .0001). No difference in lead integrity was noted between Riata and Riata ST leads (11.7% vs. 17.7% failure; P = .23). Among leads in which cable externalization was noted, laser sheaths were used more frequently (P = .01). Major complications included 3 superior vena cava/right ventricular perforations requiring surgical intervention with 1 death 12 days after the procedure and 1 pericardial effusion requiring percutaneous drainage (0.87%).

CONCLUSION

Extraction of the Riata/Riata ST leads can be challenging, and leads with externalized cables may require specific extraction techniques. Extraction of the Riata/Riata ST leads can be performed safely by experienced operators at high-volume centers with a complication rate comparable to published data.

Для цитирования: Maytin Melanie, Wilkoff Bruce L, Brunner Michael, Cronin Edmond, Love Charles J, Grazia Bongiorni Maria, Segreti Luca, Carrillo Roger G, Garisto Juan D, Kutalek Steven, Subzposh Faiz, Fischer Avi, Coffey James O, Gangireddy Sandeep R, Saba Samir, Mittal Suneet, Arshad Aysha, O'Keefe Ryan Michael, Henrikson Charles A, Belott Peter, John Roy M, Epstein Laurence M. Multicenter experience with extraction of the Riata/Riata ST ICD lead.. Heart rhythm. 2014. DOI: 10.1016/j.hrthm.2014.05.014.