DOI 10.1007/s10143-020-01381-5
A novel needle electrode for intraoperative fourth cranial nerve neurophysiological mapping.
Аннотация
OBJECTIVES
Trochlear nerve (CN-IV) mapping method has not been confirmed to date. The compound muscle action potential (CMAP) of CN-IV cannot be recorded because of the low mapping sensitivity and anatomical characteristics of the superior oblique muscle (SOM). The aim of this study was to evaluate the effectiveness of a novel needle electrode (NNE), for the intraoperative mapping of CN-IV.
MATERIALS AND
METHODS
The NNEs were inserted in the target extraocular muscles in 19 patients. We compared the CMAP amplitude of the NNE with that of the conventional needle electrode (CNE). Furthermore, we investigated the dissimilarity between the CMAP of the CN-IV and other extraocular cranial nerves (ECNs) and the correlation between the readings of the CN-IV mapping and its postoperative functional outcome.
RESULTS
The CMAP of CN-IV has been measured in nine patients (47.4%). The CMAP of CN-IV was distinguishable from other ECNs. The CMAP of the NNE was found to be three times higher than that of the CNE. Although the NNE has shown the potential to record the CN-IV's CMAP, 4 cases ended up having a CN-IV postoperative dysfunction.
CONCLUSIONS
For the first time, we confirmed the possibility of intraoperative mapping the CN-IV using an NNE inserted into the SOM. The NNE can also be useful for other neurophysiological monitoring methods.