A 58-year-old man had a bipolar epicardial left ventricular (LV) implanted for cardiac resynchronization therapy after a failed transvenous approach. The system was programmed in an LV tip-right ventricular coil configuration, but during follow-up loss of LV capture with recurrence of symptoms occurred. Changing to an LV tip-LV ring configuration restored LV pacing through anodal capture. Loss of cathodal capture was caused by dislodgment of the cathodal electrode due to a broken fixation suture. Anodal capture was used to reinstall resynchronization therapy, which resulted in clinical improvement. There were no adverse effects from anodal stimulation in a follow-up period of 6 months.

Для цитирования: Römers Hans, Bracke Frank A, Meijer Albert, Van Gelder Berry M. Preservation of ventricular capture by anodal stimulation after dislodgment of an epicardial electrode.. Pacing and clinical electrophysiology : PACE. 2006. DOI: 10.1111/j.1540-8159.2006.00288.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.