Technical note: Minimizing CIED artifacts on a 0.35 T MRI-Linac using deep learning.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Curcuru Austen N, Yang Deshan, An Hongyu, Cuculich Phillip S, Robinson Clifford G, Gach H Michael
Аннотация
BACKGROUND: Artifacts from implantable cardioverter defibrillators (ICDs) are a challenge to magnetic resonance imaging (MRI)-guided radiotherapy (MRgRT).
PURPOSE: This study tested an unsupervised generative adversarial network to mitigate ICD artifacts in balanced steady-state free precession (bSSFP) cine MRIs and improve image quality and tracking performance for MRgRT.
METHODS: Fourteen healthy volunteers (Group A) were scanned on a 0.35 T MRI-Linac with and without an MR conditional ICD taped to their left pectoral to simulate an implanted ICD. bSSFP MRI data from 12 of the volunteers were used to train a CycleGAN model to reduce ICD artifacts. The data from the remaining two volunteers were used for testing. In addition, the dataset was reorganized three times using a Leave-One-Out scheme. Tracking metrics [Dice similarity coefficient (DSC), target registration error (TRE), and 95 percentile Hausdorff distance (95% HD)] were evaluated for whole-heart contours. Image quality metrics [normalized root mean square error (nRMSE), peak signal-to-noise ratio (PSNR), and multiscale structural similarity (MS-SSIM) scores] were evaluated. The technique was also tested qualitatively on three additional ICD datasets (Group B) including a patient with an implanted ICD.
RESULTS: For the whole-heart contour with CycleGAN reconstruction: 1) Mean DSC rose from 0.910 to 0.935; 2) Mean TRE dropped from 4.488 to 2.877 mm; and 3) Mean 95% HD dropped from 10.236 to 7.700 mm. For the whole-body slice with CycleGAN reconstruction: 1) Mean nRMSE dropped from 0.644 to 0.420; 2) Mean MS-SSIM rose from 0.779 to 0.819; and 3) Mean PSNR rose from 18.744 to 22.368. The three Group B datasets evaluated qualitatively displayed a reduction in ICD artifacts in the heart.
CONCLUSION: CycleGAN-generated reconstructions significantly improved both tracking and image quality metrics when used to mitigate artifacts from ICDs.
PURPOSE: This study tested an unsupervised generative adversarial network to mitigate ICD artifacts in balanced steady-state free precession (bSSFP) cine MRIs and improve image quality and tracking performance for MRgRT.
METHODS: Fourteen healthy volunteers (Group A) were scanned on a 0.35 T MRI-Linac with and without an MR conditional ICD taped to their left pectoral to simulate an implanted ICD. bSSFP MRI data from 12 of the volunteers were used to train a CycleGAN model to reduce ICD artifacts. The data from the remaining two volunteers were used for testing. In addition, the dataset was reorganized three times using a Leave-One-Out scheme. Tracking metrics [Dice similarity coefficient (DSC), target registration error (TRE), and 95 percentile Hausdorff distance (95% HD)] were evaluated for whole-heart contours. Image quality metrics [normalized root mean square error (nRMSE), peak signal-to-noise ratio (PSNR), and multiscale structural similarity (MS-SSIM) scores] were evaluated. The technique was also tested qualitatively on three additional ICD datasets (Group B) including a patient with an implanted ICD.
RESULTS: For the whole-heart contour with CycleGAN reconstruction: 1) Mean DSC rose from 0.910 to 0.935; 2) Mean TRE dropped from 4.488 to 2.877 mm; and 3) Mean 95% HD dropped from 10.236 to 7.700 mm. For the whole-body slice with CycleGAN reconstruction: 1) Mean nRMSE dropped from 0.644 to 0.420; 2) Mean MS-SSIM rose from 0.779 to 0.819; and 3) Mean PSNR rose from 18.744 to 22.368. The three Group B datasets evaluated qualitatively displayed a reduction in ICD artifacts in the heart.
CONCLUSION: CycleGAN-generated reconstructions significantly improved both tracking and image quality metrics when used to mitigate artifacts from ICDs.