INTRODUCTION

Early infants are at risk of vitamin K (VK) deficiency-associated bleeding unless they receive VK prophylaxis. A coagulation screening test might be necessary to detect prophylactic failure, but it is rarely conducted owing to the difficulty of blood sampling. CoaguChek XS is a portable, point-of-care (POC) coagulation analyzer that determines international normalized ratios (INRs) using only small volumes of blood. Although POC tests assessing INR have become widespread for adults, studies have not been performed in early infancy. This study aimed to determine the reference interval (RI) for INR for 1-month-old infants using the CoaguChek XS and to compare INRs with Normotest (NT) values to evaluate its efficacy as a VK deficiency screening test.

MATERIALS AND

METHODS

This study included 578 consecutive 1-month-old, healthy, term infants who underwent a 1-month health checkup. CoaguChek XS INRs and NT values were analyzed for each subject.

RESULTS

The enrolled neonates were orally administered VK 6-12h and 4days after birth. An INR RI for 1-month-old infants was established using the CoaguChek XS (range of 0.9-1.1). On the other hand, the NT RI was 69%-130%. We observed a significant inverse correlation between INRs and NT values.

CONCLUSIONS

Our study is the first to establish an INR RI using the CoaguChek XS for 1-month-old, healthy, term infants. CoaguChek XS INR was more convenient to test than NT. However, we could not demonstrate its efficacy as a VK deficiency screening test because no infant had an aberrant NT value or a bleeding disorder.

Для цитирования: Iijima Shigeo, Baba Toru, Ueno Daizo, Ohishi Akira. International normalized ratio testing with a point-of-care coagulometer in 1-month-old infants: A comparison with Normotest.. Thrombosis research. 2016. DOI: 10.1016/j.thromres.2016.08.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.