The Annals of otology, rhinology, and laryngology · 2021
Electrostatic Discharge (ESD) as a Cause of Facial Nerve Stimulation After Cochlear Implantation: A Case Report.
Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rath Smruti, Glaun Mica, Emery Claudia, Liu Yi-Chun Carol
Аннотация
OBJECTIVE: To discuss persistent facial nerve stimulation (FNS) related to repeated electrostatic discharge (ESD) shock following cochlear implantation.
METHODS: Single case report with literature review.
RESULTS: FNS is a feared complication after cochlear implantation, occurring in approximately 7% of cases, with most patients having anatomic abnormalities. The presented case has no anatomical abnormalities but reported frequent environmental static shock. FNS during the first 1 to 3 seconds of processor attachment caused a significant decrease in the patient's quality of life, requiring subsequent re-implantation with full resolution.
CONCLUSIONS: FNS is a complication of cochlear implantation that can cause a great deal of distress and discomfort. Frequent electrostatic discharge (ESD) contributed to device malfunctioning and FNS in a patient with otherwise normal anatomy and should be avoided if possible.
METHODS: Single case report with literature review.
RESULTS: FNS is a feared complication after cochlear implantation, occurring in approximately 7% of cases, with most patients having anatomic abnormalities. The presented case has no anatomical abnormalities but reported frequent environmental static shock. FNS during the first 1 to 3 seconds of processor attachment caused a significant decrease in the patient's quality of life, requiring subsequent re-implantation with full resolution.
CONCLUSIONS: FNS is a complication of cochlear implantation that can cause a great deal of distress and discomfort. Frequent electrostatic discharge (ESD) contributed to device malfunctioning and FNS in a patient with otherwise normal anatomy and should be avoided if possible.