Electrostatic Discharge (ESD) as a Cause of Facial Nerve Stimulation After Cochlear Implantation: A Case Report.
Аннотация
OBJECTIVE
To discuss persistent facial nerve stimulation (FNS) related to repeated electrostatic discharge (ESD) shock following cochlear implantation.
METHODS
Single case report with literature review.
RESULTS
FNS is a feared complication after cochlear implantation, occurring in approximately 7% of cases, with most patients having anatomic abnormalities. The presented case has no anatomical abnormalities but reported frequent environmental static shock. FNS during the first 1 to 3 seconds of processor attachment caused a significant decrease in the patient's quality of life, requiring subsequent re-implantation with full resolution.
CONCLUSIONS
FNS is a complication of cochlear implantation that can cause a great deal of distress and discomfort. Frequent electrostatic discharge (ESD) contributed to device malfunctioning and FNS in a patient with otherwise normal anatomy and should be avoided if possible.