Intraoperative lateral rectus electromyographic recordings optimized by deep intraorbital needle electrodes.

Oishi Tatsuya, Van Gompel Jamie J, Link Michael J, Tooley Andrea A, Hoffman Ernest M

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

We demonstrate the advantages and safety of long, intraorbitally-placed needle electrodes, compared to standard-length subdermal electrodes, when recording lateral rectus electromyography (EMG) during intracranial surgeries.

METHODS

Insulated 25 mm and uninsulated 13 mm needle electrodes, aimed at the lateral rectus muscle, were placed in parallel during 10 intracranial surgeries, examining spontaneous and stimulation-induced EMG activities. Postoperative complications in these patients were reviewed, alongside additional patients who underwent long electrode placement in the lateral rectus.

RESULTS

In 40 stimulation-induced recordings from 10 patients, the 25 mm electrodes recorded 6- to 26-fold greater amplitude EMG waveforms than the 13 mm electrodes. The 13 mm electrodes detected greater unwanted volume conduction upon facial nerve stimulation, typically exceeding the amplitude of abducens nerve stimulation. Except for one case with lateral canthus ecchymosis, no clinical or radiographic complications occurred in 36 patients (41 lateral rectus muscles) following needle placement.

CONCLUSIONS

Intramuscular recordings from long electrode in the lateral rectus offers more reliable EMG monitoring than 13 mm needles, with excellent discrimination between abducens and facial nerve stimulations, and without significant complications from needle placement.

SIGNIFICANCE: Long intramuscular electrode within the orbit for lateral rectus EMG recording is practical and reliable for abducens nerve monitoring.

Для цитирования: Oishi Tatsuya, Van Gompel Jamie J, Link Michael J, Tooley Andrea A, Hoffman Ernest M. Intraoperative lateral rectus electromyographic recordings optimized by deep intraorbital needle electrodes.. Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology. 2021. DOI: 10.1016/j.clinph.2021.08.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.