SmartPulse, a machine learning approach for calibration-free dynamic RF shimming: Preliminary study in a clinical environment.

Tomi-Tricot Raphaël, Gras Vincent, Thirion Bertrand, Mauconduit Franck, Boulant Nicolas, Cherkaoui Hamza, Zerbib Pierre, Vignaud Alexandre, Luciani Alain, Amadon Alexis

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: A calibration-free pulse design method is introduced to alleviate B1+ artifacts in clinical routine with parallel transmission at high field, dealing with significant inter-subject variability, found for instance in the abdomen.

THEORY AND

METHODS

From a dual-transmit 3T scanner, a database of B1+ and off-resonance abdominal maps from 50 subjects was first divided into 3 clusters based on mutual affinity between their respective tailored kT -points pulses. For each cluster, a kT -points pulse was computed, minimizing normalized root-mean-square flip angle deviations simultaneously for all subjects comprised in it. Using 30 additional subjects' field distributions, a machine learning classifier was trained on this 80-labeled-subject database to recognize the best pulse from the 3 ones available, relying only on patient features accessible from the preliminary localizer sequence present in all protocols. This so-called SmartPulse process was experimentally tested on an additional 53-subject set and compared with other pulse types: vendor's hard calibration-free dual excitation, tailored static radiofrequency shimming, universal and tailored kT -points pulses.

RESULTS

SmartPulse outperformed both calibration-free approaches. Tailored static radiofrequency shimming yielded similar flip angle homogeneity for most patients but broke down for some while SmartPulse remained robust. Although flip angle homogeneity was systematically better with tailored kT -points, the difference was barely noticeable on in vivo images.

CONCLUSION

The proposed method paves the way toward an efficient trade-off between tailored and universal pulse design approaches for large inter-subject variability. With no need for on-line field mapping or pulse design, it can fit seamlessly into a clinical protocol.

Для цитирования: Tomi-Tricot Raphaël, Gras Vincent, Thirion Bertrand, Mauconduit Franck, Boulant Nicolas, Cherkaoui Hamza, Zerbib Pierre, Vignaud Alexandre, Luciani Alain, Amadon Alexis. SmartPulse, a machine learning approach for calibration-free dynamic RF shimming: Preliminary study in a clinical environment.. Magnetic resonance in medicine. 2019. DOI: 10.1002/mrm.27870.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.