MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.prosdent.2024.07.032

Медицинская библиотека · The Journal of prosthetic dentistry · 2024

Accuracy of an autonomous dental implant robotic system in placing tilted implants for edentulous arches.

Xi Shuang, Hu Jun, Yue Guangna, Wang Shaohai

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

STATEMENT OF PROBLEM: Accurate placement of tilted implants is essential as they are typically close to important anatomic structures. Inaccurate implant position may damage those structures and affect outcomes.

PURPOSE: The purpose of this in vitro study was to compare the accuracy and efficiency of an autonomous dental implant robotic (ADIR) system and a static computer-assisted implant surgery (sCAIS) system in placing tilted implants in edentulous patients.

MATERIAL AND

METHODS

Ten 3-dimensionally (3D) printed edentulous mandibular casts were assigned to 1 of 2 groups (ADIR and sCAIS). The coronal, apical, and angular deviations of the placement of tilted implants, preoperative preparation time, and surgical time were compared between the 2 groups. The paired samples t test and the independent samples t test were used to compare the groups (α=.05).

RESULTS

The mean ±standard deviation of coronal, apical, and angular deviation in the ADIR group and sCAIS group were 0.47 ±0.06 mm versus 1.09 ±0.11 mm, 0.47 ±0.05 mm versus 1.53 ±0.14 mm, and 0.91 ±0.82 degrees versus 2.83 ±0.55 degrees, respectively. The deviations of the tilted implant positions in the ADIR were relatively small and significantly different from those of sCAIS (P<.05). The preoperative preparation time of the ADIR group was significantly longer than that of the sCAIS group (P<.001), and the surgical time for the 2 groups was statistically similar (P=.259).

CONCLUSIONS

Compared with the sCAIS system, the deviation of tilted implants in the ADIR group was smaller, but the preoperative preparation time was longer. The results indicated that using the ADIR for tilted implantation can lead to more accurate implantation positions and reduce the occurrence of complications. However, it is time consuming, and the workflow should be simplified to improve efficiency.

Для цитирования: Xi Shuang, Hu Jun, Yue Guangna, Wang Shaohai. Accuracy of an autonomous dental implant robotic system in placing tilted implants for edentulous arches.. The Journal of prosthetic dentistry. 2024. DOI: 10.1016/j.prosdent.2024.07.032.