[Assessment of the clinical feasibility of OptiGo for hand-held echocardiography].

Kawai Junichi, Tanabe Kazuaki, Yagi Toshikazu, Fujii Yoko, Konda Toshiko, Sumida Toshiaki, Okada Midori, Yamaguchi Kazuto, Tani Tomoko, Yamabe Kenji, Morioka Shigefumi

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Hand-held ultrasound devices are becoming available for clinical examination, but the accuracy and precision of such devices are unclear. This study compared the accuracy of a hand-held echo device to a standard echo system.

METHODS

Twenty-two patients were examined with the OptiGo (Phillips Medial Systems) hand-held ultrasound system with a 2.5 MHz transducer and SONOS 5500 (Phillips Medial Systems) standard ultrasound system with a 2 to 4 MHz wideband transducer. Patients with cardiac arrhythmia and tachycardia were excluded. Image quality, chamber size (left ventricle, left atrium), global and regional left ventricular function, valve morphology and severity of valve regurgitation were assessed.

RESULTS

There was good agreement between the two imaging devices for image quality (77.3%), left ventricular ejection fraction (90.5%), regional wall motion score (> 90%), valve morphology (> 90%), severity of valve regurgitation (> 81.0%) and there was good correlation and agreement for left ventricular and left atrial size.

CONCLUSIONS

Although the OptiGo has limitations, the hand-held examination appropriately estimates global and regional left ventricular function, valve morphology, valve regurgitation and chamber size.

Для цитирования: Kawai Junichi, Tanabe Kazuaki, Yagi Toshikazu, Fujii Yoko, Konda Toshiko, Sumida Toshiaki, Okada Midori, Yamaguchi Kazuto, Tani Tomoko, Yamabe Kenji, Morioka Shigefumi. [Assessment of the clinical feasibility of OptiGo for hand-held echocardiography].. Journal of cardiology. 2003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.