Feasibility study of an automated Strabismus screening Test using Augmented Reality and Eye-tracking (STARE).

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Nixon Nisha, Thomas Peter B M, Jones Pete R

Аннотация

BACKGROUND: New digital technologies (augmented reality headsets, eye-tracking) may potentially allow for automated assessments of ocular misalignment. Here, we evaluate the feasibility of a novel, open-source strabismus test ("STARE") as an automated screening tool.

METHODS: Work progressed in 2 phases. In phase 1 ("development"), we used Fresnel prisms to elicit horizontal misalignments of known magnitude (1-40 prism dioptres) in orthotropic controls. In phase 2 ("validation"), we applied the system to adults with an established diagnosis of strabismus, and quantified the ability of the test to distinguish between those with horizontal misalignment and those without. Agreement between the alternate prism cover test measurements and STARE measurements was computed using Bland-Altman plots and product-moment correlation coefficients.

RESULTS: Seven orthotropic controls and nineteen patients with strabismus were recruited (mean age 58.7 ± 22.4 years). STARE was able to identify the presence of horizontal strabismus with an area under the curve of 1.00 (100% sensitivity and 100% specificity). The mean difference (bias) {95% CI} was 2.1 {-1.8, 9.9} prism dioptres, and the 95% coefficient of repeatability {95% CI} was ±27.9 {14.8, 50.8} prism dioptres. The Pearson correlation between APCT and STARE was r24 = 0.62, P < 0.001.

CONCLUSIONS: STARE shows promise as a simple, automated tool for performing a screening assessment of strabismus. It is a rapid (60 s) test that can be performed using a consumer augmented reality headset with integrated eye-tracking, and might conceivably be used remotely by non-specialists in future as a means of highlighting individuals needing face-to-face specialist care.