Factors Affecting the Irrigation Fluid Temperature During Laser Lithotripsy: In Vitro Experimental Study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Peteinaris Angelis, Tsaturyan Arman, Pantazis Lampros, Farsari Ergina, Ballesta Martinez Begoa, Pagonis Konstantinos, Adamou Constantinos, Vagionis Athanasios, Natsos Anastasios, Liatsikos Evangelos, Kallidonis Panagiotis
Аннотация
OBJECTIVE: To investigate the effect of the diameter of laser fiber, pelvis volume, presence and type of the stone on irrigation fluid temperature rise.
MATERIAL AND METHODS: A 20ml syringe, 12/14 ureteral access sheath(UAS), a dual-lumen catheter and a thermocouple were used. The 12/14Fr UAS(Cook Ireland Ltd., Limerick, Ireland) and the Thermocouple(SE001, Pico Technologies, Cambridgeshire, UK) were inserted in the syringe. The syringe was closed allowing outflow from the UAS with rate at 10ml/min. The Quanta Ho 150W(Quanta System, Samarate, Italy) laser was used and fired with 10W(2Jx5Hz), 20W(2 × 10 Hz), 40W(2 × 20 Hz), 60W(2 × 30 Hz). These power settings were tested in different conditions: fibers(200µm, 365µm and 550µm), volumes(5ml, 10ml and 20ml) and artificial stones(soft, hard). The laser was activated for 30 seconds and reactivation was performed when the temperature reached below 26 °C.
RESULTS: For all trials 60W of energy resulted in higher temperature rise. No differences were observed when different fibers were used. The highest temperatures (up to 80 °C) for 60W were reported in 5ml syringe and the lowest (<45 °C) with 20ml. The maximal temperature of >59°C was recorded for the power of 60W(1Jx60Hz). The temperature exceeded 43 °C when power settings >40W were applied.
CONCLUSION: Increasing the overall power, increases the irrigation fluid temperature significantly. The smaller the volume of the pelvis, the greater the temperature elevation. The fiber size did not affect the temperature increase pattern. The presence of artificial stones was associated with the absorption of energy emitted by the laser.
MATERIAL AND METHODS: A 20ml syringe, 12/14 ureteral access sheath(UAS), a dual-lumen catheter and a thermocouple were used. The 12/14Fr UAS(Cook Ireland Ltd., Limerick, Ireland) and the Thermocouple(SE001, Pico Technologies, Cambridgeshire, UK) were inserted in the syringe. The syringe was closed allowing outflow from the UAS with rate at 10ml/min. The Quanta Ho 150W(Quanta System, Samarate, Italy) laser was used and fired with 10W(2Jx5Hz), 20W(2 × 10 Hz), 40W(2 × 20 Hz), 60W(2 × 30 Hz). These power settings were tested in different conditions: fibers(200µm, 365µm and 550µm), volumes(5ml, 10ml and 20ml) and artificial stones(soft, hard). The laser was activated for 30 seconds and reactivation was performed when the temperature reached below 26 °C.
RESULTS: For all trials 60W of energy resulted in higher temperature rise. No differences were observed when different fibers were used. The highest temperatures (up to 80 °C) for 60W were reported in 5ml syringe and the lowest (<45 °C) with 20ml. The maximal temperature of >59°C was recorded for the power of 60W(1Jx60Hz). The temperature exceeded 43 °C when power settings >40W were applied.
CONCLUSION: Increasing the overall power, increases the irrigation fluid temperature significantly. The smaller the volume of the pelvis, the greater the temperature elevation. The fiber size did not affect the temperature increase pattern. The presence of artificial stones was associated with the absorption of energy emitted by the laser.