European journal of paediatric dentistry · 2014
Accuracy of three different apex locators and visual exam in primary teeth with and without root resorption in vitro.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Oznurhan F, Tüzüner T, Baygin O, Unal M, Kapdan A, Ozturk C
Аннотация
AIM: The aim of this study was to evaluate the accuracy and reliability of three apex locators (EndoMaster, Raypex, DentaPort ZX) and visual assessment in primary molar teeth in vitro.
MATERIALS AND METHODS: Twenty-four extracted human primary molar teeth with and without root resorption were used in this study. After endodontic access preparation, root canal length was visually measured by means of a K file; then the teeth were embedded into alginate and the roots were measured again by mans of the three apex locators. The results were collected in SPSS 15.0 and statistical evaluations were completed by one-way ANOVA and Kruksal-Wallis test.
RESULTS: The measurements closest to the visually determined length were obtained with EndoMaster. No statistically significant differences were found between visual and apex locators lengths (p>0.05).
CONCLUSION: Within the limitations of this study, the use of apex locators would be useful in the endodontic treatment of primary teeth.
MATERIALS AND METHODS: Twenty-four extracted human primary molar teeth with and without root resorption were used in this study. After endodontic access preparation, root canal length was visually measured by means of a K file; then the teeth were embedded into alginate and the roots were measured again by mans of the three apex locators. The results were collected in SPSS 15.0 and statistical evaluations were completed by one-way ANOVA and Kruksal-Wallis test.
RESULTS: The measurements closest to the visually determined length were obtained with EndoMaster. No statistically significant differences were found between visual and apex locators lengths (p>0.05).
CONCLUSION: Within the limitations of this study, the use of apex locators would be useful in the endodontic treatment of primary teeth.