BACKGROUND AND AIM OF THE STUDY: To investigate the accuracy of two methods of measuring features in cardiac anatomy, using an objective standard cast model.

METHODS

We made a silicone cast using a swine heart. Computerized tomography data of the solidified cast were processed through virtual reality (VR) software and through two-dimensional multiplanar-reconstruction (2D-MPR), and all measurements were compared against physical measurements of the cast.

RESULTS

The cast perfectly demonstrated the fine detail of the aortic valve and the proximal parts of coronary arteries. Anatomical features were measured by 3D-VR, 2D-MPR, and directly on the cast. Measurement differences between 2D-MPR and the cast were on average at least 3.6 times larger than those between 3D-VR and the cast.

CONCLUSIONS

Based on the observed accuracy, 3D-VR measurements seem considerably more accurate than the current standard 2D-MPR, and 3D-VR may be considered as the next gold standard for 3D measurement of cardiac anatomy in vivo.

Для цитирования: Kamiya Kenichi, Matsubayashi Yuji, Terada Shinya, Nagatani Yukihiro, Fujii Taihei, Nakata Susumu, Suzuki Tomoaki. Ex-vivo aortic root and coronary artery cast measurement to validate the accuracy of virtual imaging.. Journal of cardiac surgery. 2022. DOI: 10.1111/jocs.16559.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.