Evaluating the performance of automated UV enzymatic assay for screening of glucose 6-phosphate dehydrogenase deficiency.

Клиническое исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Anantasomboon Pornchai, Chanda Makamas, Jugnam-Ang Watcharapong, Witoonpanich Pirada, Cheepsunthorn Poonlarp, Nuchprayoon Issarang, Fucharoen Suthat, Cheepsunthorn Chalisa L

Аннотация

INTRODUCTION: A precise and reliable screening assay for glucose 6-phosphate dehydrogenase (G6PD) deficiency would greatly help avoiding unwanted outcomes due to bilirubin neurotoxicity in neonatal jaundice and antimalarial-induced haemolytic anaemia in malaria patients. Currently, available assays are laborious and require sophisticated laboratory expertise. This study aimed to evaluate the performance of a recently introduced automated screening assay for G6PD deficiency by comparing with a routine spectrophotometric assay.

METHODS: An automated UV-based enzymatic (Mindray, PRC) and spectrophotometric assays were performed simultaneously in parallel to determine G6PD activity in 251 blood samples from the subjects.

RESULTS: The median G6PD activity value from spectrophotometric assay was significantly lower than that of from the automated assay. The mean difference was -2.0 U/g haemoglobin (-7.3 to 3.2; P < 0.0001). The mean activity values of both assays were strongly correlated with Pearson's correlation coefficient of r = 0.8. Cohen's kappa statistics between assays was 0.77 (0.70-0.83). The sensitivity, specificity, positive and negative predictive values of the automated assay were 85.7%, 99.2%, 85.7%, 99.2%, respectively. The sensitivity and positive predictive values of the automated assay for identifying intermediate G6PD activity levels were 40.0% and 25.0%, respectively. Genotyping was performed to confirm G6PD deficient and intermediate samples. The turnaround time for 40 samples was 60 minutes for the automated assay and 300 minutes for spectrophotometric assay.

CONCLUSION: The automated assay for the detection of G6PD deficiency is comparable to a routine spectrophotometric assay and help reducing sample handling time. However, the assay shows limitation in identifying individuals with G6PD intermediate.