A device to measure the cross-sectional area of soft connective tissues.

Vanderby R, Masters G P, Bowers J R, Graf B K

СтатьяПеревод пока не подготовлен

Аннотация

A device has been designed and fabricated to measure the cross-sectional area of soft connective tissues ex vivo. It consists of two displacement transducers; one sensing tissue thickness and the other sensing width. Outputs are recorded (via an analog to digital interface) using a personal computer. Numerical integration of a thickness versus width plot computes cross-sectional area. This plot also provides a quality check of acquired data. This device has been successfully used in biomechanical studies of rabbit patellar tendons, rat medial collateral ligaments, and dissected specimens of human fascia.

Для цитирования: Vanderby R, Masters G P, Bowers J R, Graf B K. A device to measure the cross-sectional area of soft connective tissues.. IEEE transactions on bio-medical engineering. 1991. DOI: 10.1109/10.88449.

Предметные рубрики MeSH