DOI 10.1590/2175-8239-jbn-2019-0236
Concordance analysis between dosed serum bicarbonate and that calculated by gas analysis in chronic renal patients.
Аннотация
INTRODUCTION
The control of metabolic acidosis in dialysis patients focuses on the supply of bicarbonate during the dialysis session, and it is not standard in all hemodialysis to assess serum bicarbonate concentrations. Bicarbonate expressed in blood gas analysis is the most sensitive standard of analysis and it is measured indirectly, using the Henderson-Hasselbalch equation. There are no studies in this population evaluating the concordance between the calculated bicarbonate with the direct method of biochemical analysis. The aim of this study was to analyze the concordance between the measured and calculated serum bicarbonate levels using blood gas analysis.
METHODS
We analyzed blood samples from chronic kidney patients undergoing hemodialysis, using the same sample of bicarbonate analysis by biochemistry and gasometry. The concordance was assessed using the Bland-Altman method.
RESULTS
51 samples were analyzed. The analysis revealed a high correlation (r = 0.73) and a mean difference (bias) of 1.15 ± 3 mmol/L. The median time between collection and examination was 241 minutes.
DISCUSSION: We can conclude that the biochemical bicarbonate analysis compared to that calculated from blood gas analysis in chronic renal patients was consistent. For greater concordance between the data, it is important that the time between the collection of the samples and the referral to the laboratory for carrying out the dosages does not exceed four hours. The serum bicarbonate dosage can result in cost savings when compared to that of bicarbonate in blood gas analysis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.