Cardiovascular screening in low-income settings using a novel 4-lead smartphone-based electrocardiograph (D-Heart®).

Maurizi Niccolo', Faragli Alessandro, Imberti Jacopo, Briante Nicolò, Targetti Mattia, Baldini Katia, Sall Amadou, Cisse Abibou, Berzolari Francesca Gigli, Borrelli Paola, Avvantaggiato Fulvio, Perlini Stefano, Marchionni Niccolo', Cecchi Franco, Parigi Gianbattista, Olivotto Iacopo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

MHealth technologies are revolutionizing cardiovascular medicine. However, a low-cost, user-friendly smartphone-based electrocardiograph is still lacking. D-Heart® is a portable device that enables the acquisition of the ECG on multiple leads which streams via Bluetooth to any smartphone. Because of the potential impact of this technology in low-income settings, we determined the accuracy of D-Heart® tracings in the stratification of ECG morphological abnormalities, compared with 12-lead ECGs.

METHODS

Consecutive African patients referred to the Ziguinchor Regional Hospital (Senegal) were enrolled (n=117; 69 males, age 39±11years). D-Heart® recordings (3 peripheral leads plus V5) were obtained immediately followed by 12 lead ECGs and were assessed blindly by 2 independent observers. Global burden of ECG abnormalities was defined by a semi-quantitative score based on the sum of 9 criteria, identifying four classes of increasing severity.

RESULTS

D-Heart® and 12-lead ECG tracings were respectively classified as: normal: 72 (61%) vs 69 (59%); mildly abnormal: 42 (36%) vs 45 (38%); moderately abnormal: 3 (3%) vs 3 (3%). None had markedly abnormal tracings. Cohen's weighted kappa (kw) test demonstrated a concordance of 0,952 (p<0,001, agreement 98,72%). Concordance was high as well for the Romhilt-Estes score (kw=0,893; p<0,001 agreement 97,35%). PR and QRS intervals comparison with Bland-Altman method showed good accuracy for D-Heart® measurements (95% limit of agreement ±20ms for PR and ±10ms for QRS).

CONCLUSIONS

D-Heart® proved effective and accurate stratification of ECG abnormalities comparable to the 12-lead electrocardiographs, thereby opening new perspectives for low-cost community cardiovascular screening programs in low-income settings.

Для цитирования: Maurizi Niccolo', Faragli Alessandro, Imberti Jacopo, Briante Nicolò, Targetti Mattia, Baldini Katia, Sall Amadou, Cisse Abibou, Berzolari Francesca Gigli, Borrelli Paola, Avvantaggiato Fulvio, Perlini Stefano, Marchionni Niccolo', Cecchi Franco, Parigi Gianbattista, Olivotto Iacopo. Cardiovascular screening in low-income settings using a novel 4-lead smartphone-based electrocardiograph (D-Heart®).. International journal of cardiology. 2017. DOI: 10.1016/j.ijcard.2017.02.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.