DOI 10.1016/j.clinbiomech.2015.08.016

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2015

Optimal internal fixation of anatomically shaped synthetic bone grafts for massive segmental defects of long bones.

Vorys George C, Bai Hanying, Chandhanayingyong Chandhanarat, Lee Chang H, Compton Jocelyn T, Caldwell Jon-Michael, Gardner Thomas R, Mao Jeremy J, Lee Francis Y

СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Large segmental bone defects following tumor resection, high-energy civilian trauma, and military blast injuries present significant clinical challenges. Tissue engineering strategies using scaffolds are being considered as a treatment, but there is little research into optimal fixation of such scaffolds.

METHODS

Twelve fresh-frozen paired cadaveric legs were utilized to simulate a critical sized intercalary defect in the tibia. Poly-ε-caprolactone and hydroxyapatite composite scaffolds 5 cm in length with a geometry representative of the mid-diaphysis of an adult human tibia were fabricated, inserted into a tibial mid-diaphyseal intercalary defect, and fixed with a 14-hole large fragment plate. Optimal screw fixation comparing non-locking and locking screws was tested in axial compression, bending, and torsion in a non-destructive manner. A cyclic torsional test to failure under torque control was then performed.

FINDINGS: Biomechanical testing showed no significant difference for bending or axial stiffness with non-locking vs. locking fixation. Torsional stiffness was significantly higher (P=0.002) with the scaffold present for both non-locking and locking compared to the scaffold absent. In testing to failure, angular rotation was greater for the non-locking compared to locking constructs at each torque level up to 40 N-m (P<0.05). The locking constructs survived a significantly higher number of loading cycles before reaching clinical failure at 30 degrees of angular rotation (P<0.02).

INTERPRETATION: The presence of the scaffold increased the torsional stiffness of the construct. Locking fixation resulted in a stronger construct with increased cycles to failure compared to non-locking fixation.

Для цитирования: Vorys George C, Bai Hanying, Chandhanayingyong Chandhanarat, Lee Chang H, Compton Jocelyn T, Caldwell Jon-Michael, Gardner Thomas R, Mao Jeremy J, Lee Francis Y. Optimal internal fixation of anatomically shaped synthetic bone grafts for massive segmental defects of long bones.. Clinical biomechanics (Bristol, Avon). 2015. DOI: 10.1016/j.clinbiomech.2015.08.016.

Предметные рубрики MeSH