Evaluation of the suitability of the World Health Organization International Reference Reagent for Hb A2 quantitation (89/666) for continued use.

De la Salle Barbara, Stephens Adrian D, Wild Barbara J, Harteveld Cornelis L, Hyde Keith

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

The accurate determination of Hb A2 is a key marker when screening for a β-thalassaemia carrier. Data from external quality assessment (EQA) exercises have shown a lack of alignment of Hb A2 quantitation both within and between methods. The only reference material available for Hb A2 quantitative assay at the time of writing is the World Health Organization International Reference Reagent (89/666; WHO IRR) prepared in the 1980s and not validated for all current methodologies.

METHOD: The WHO IRR was analysed for Hb A2 concentration by 52 laboratories using a representative range of high-performance liquid chromatography and capillary electrophoresis analysers. The results of the analysis were compared to those of a whole blood EQA specimen of similar Hb A2 concentration, distributed in the same week.

RESULTS

The mean Hb A2 value obtained for the WHO IRR was 5.17%, compared to the assigned value of 5.3%. The range of results returned was wide (4.0%-6.2%), with differences in the results observed by between and within analyser groups. A similar range of results was seen with the whole blood sample, although the bias observed between analyser types was different from that seen with the WHO IRR.

CONCLUSION

The results may indicate a lack of commutability of the WHO IRR material, resulting from deterioration, matrix effects or changes in reagent formulation or calibration parameters. Further examination of the suitability of the WHO IRR (89/666) for continued use is required.

Для цитирования: De la Salle Barbara, Stephens Adrian D, Wild Barbara J, Harteveld Cornelis L, Hyde Keith. Evaluation of the suitability of the World Health Organization International Reference Reagent for Hb A2 quantitation (89/666) for continued use.. International journal of laboratory hematology. 2019. DOI: 10.1111/ijlh.12966.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.