MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1253/circj.cj-14-1113

Медицинская библиотека · Circulation journal : official journal of the Japanese Circulation Society · 2015

In vivo assessment of novel stentless valve in the mitral position.

Kainuma Satoshi, Kasegawa Hitoshi, Miyagawa Shigeru, Nishi Hiroyuki, Yaku Hitoshi, Takanashi Shuichiro, Hashimoto Kazuhiro, Okada Yukikatsu, Nakatani Satoshi, Umezu Mitsuo, Daimon Takashi, Sakaguchi Taichi, Toda Koichi, Sawa Yoshiki, Committee of the Japanese Society of Stentless Mitral Valve

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

We conducted in vivo examinations of a newly designed stentless mitral valve (SMV), formed by suturing 2 leaflets with the "legs" serving as chorda tendinea, made from bovine pericardium, to a flexible ring. METHODS AND

RESULTS

Seven pigs underwent implantation of the SMV constructed with a 23-mm (n=5) or 25-mm (n=2) Duran ring. Baseline echocardiography examinations were used to evaluate the annular anteroposterior diameter, and distance between the mitral annulus (MA) and papillary muscles (PMs) to determine SMV-leg length. After removing the native valve, the SMV-legs were fixed to the anterior and posterior PMs, followed by fixation of the ring to the native MA. Immediately after surgery, all animals presented none or trivial mitral regurgitation, with mean and peak trans-SMV pressure gradient values of 1.9±0.8 and 6.0±3.1 mmHg, respectively. The mean length of the SMV-leg was 19.4±3.9 mm, which correlated with the distance between anterior and posterior MA-PM (r=0.96 and 0.94, respectively, P<0.01 for both). The discrepancy between the anteroposterior diameter of the ring (outside diameter) and that of the native valve was 1.0±2.9 mm, which correlated with the trans-SMV pressure gradient (r=0.81, P=0.025).

CONCLUSIONS

In our preliminary study, the SMV demonstrated excellent diastolic inflow dynamics and closing function in vivo. Preoperative precise assessment of MV configuration may serve as a basis for selection of appropriate ring size and SMV-leg length.

Для цитирования: Kainuma Satoshi, Kasegawa Hitoshi, Miyagawa Shigeru, Nishi Hiroyuki, Yaku Hitoshi, Takanashi Shuichiro, Hashimoto Kazuhiro, Okada Yukikatsu, Nakatani Satoshi, Umezu Mitsuo, Daimon Takashi, Sakaguchi Taichi, Toda Koichi, Sawa Yoshiki, Committee of the Japanese Society of Stentless Mitral Valve. In vivo assessment of novel stentless valve in the mitral position.. Circulation journal : official journal of the Japanese Circulation Society. 2015. DOI: 10.1253/circj.cj-14-1113.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.