DOI 10.1007/s00405-024-08854-z
Safety and efficacy of the mAXIS stapes prosthesis.
Аннотация
PURPOSE: Otosclerosis leads to a fixed stapes footplate and thus to hearing loss. The predominant treatment method is surgery, with various types of stapes prostheses available. The aim of this study was to investigate the safety and efficacy of the new mAXIS Stapes Prosthesis.
METHODS
34 cases of otosclerosis were implanted with the new mAXIS Stapes Prosthesis. Comprehensive clinical assessments, including pre- and postoperative pure tone audiometry was performed at short-term (ST) follow-up at 25 (± 15) days and mid-term (MT) follow-up at 181 (± 107) days. The pure tone average of 0.5, 1, 2 and 3 kHz (PTA4) was calculated.
RESULTS
In all cases, the application of the prosthesis was successful and straightforward. The postoperative PTA4 air-bone gap was 10.7 ± 5.2 dB at ST follow-up (n = 34) and 8.3 ± 4.1 dB at MT follow-up (n = 18). In 61% of cases, the ABG-closure was within 10 dB and in 100% of cases within 20 dB at MT follow-up.
CONCLUSION
Findings of this study support that the mAXIS Stapes Prosthesis is safe for implantation and shows promising audiological outcome. Future investigations will contribute its long-term efficacy and safety profile.