Bioprosthetic valve fracture: Predictors of outcome and follow-up. Results from a multicenter study.

Brinkmann Christina, Abdel-Wahab Mohamed, Bedogni Francesco, Bhadra Oliver D, Charbonnier Gaetan, Conradi Lenard, Hildick-Smith David, Kargoli Faraj, Latib Azeem, Van Mieghem Nicolas M, Miura Mizuki, Mylotte Darren, Landes Uri, Pilgrim Thomas, Riess Friedrich-Christian, Taramasso Maurizio, Tchétché Didier, Testa Luca, Thiele Holger, Webb John, Windecker Stephan, Witt Julian, Wohlmuth Peter, Wolf Alexander, Schofer Joachim

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

To evaluate outcome and its predictors of bioprosthetic valve fracture (BVF) in patients undergoing valve-in-valve transcatheter aortic valve replacement (VIV-TAVR).

BACKGROUND

BVF is feasible and reduces transvalvular gradients in VIV-TAVR-procedures, but follow-up-data and information on factors influencing the outcome are missing.

METHODS

The 81 cases of BVF-VIV-TAVR were collected from 14 international centers.

RESULTS

Predominantly transcatheter heart valve (THV) was implanted first, followed by BVF. VARC-2 defined device success was 93%, most failures were attributed to residual high gradients. Mean gradients decreased from 37 ± 13 mmHg to 10.8 ± 5.9 mmHg (p < 0.001). BVF reduced the gradient by 16 mmHg. During follow-up (FU, 281 ± 164 days) mean gradient remained stable (10.8 ± 5.9 mmHg at discharge, 12.4 ± 6.3 mmHg at FU, p = ns). In-hospital major adverse events occurred in 3.7%. Event-free survival at 276 ± 237.6 days was 95.4%. The linear mixed model identified balloon-expandable valves (BEV), Mitroflow surgical valve, stenotic surgical bioprostheses and balloon only 1 mm larger than the true internal diameter of the surgical valve as predictors for higher gradients.

CONCLUSIONS

BVF is safe and can significantly reduce gradients, which remain stable at FU. BEV, Mitroflow surgical valve, stenotic bioprostheses and balloon larger than the true internal diameter of the surgical valve of only 1 mm are predictors for higher final gradients.

Для цитирования: Brinkmann Christina, Abdel-Wahab Mohamed, Bedogni Francesco, Bhadra Oliver D, Charbonnier Gaetan, Conradi Lenard, Hildick-Smith David, Kargoli Faraj, Latib Azeem, Van Mieghem Nicolas M, Miura Mizuki, Mylotte Darren, Landes Uri, Pilgrim Thomas, Riess Friedrich-Christian, Taramasso Maurizio, Tchétché Didier, Testa Luca, Thiele Holger, Webb John, Windecker Stephan, Witt Julian, Wohlmuth Peter, Wolf Alexander, Schofer Joachim. Bioprosthetic valve fracture: Predictors of outcome and follow-up. Results from a multicenter study.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29755.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.